All metals, with the exception of high noble metals, are susceptible to corrosion. This problem is exacerbated in the body, where implants must function in aqueous environments under complex loading and often utilize designs that contain inherent stress concentrations and material discontinuities. How would you design a modular hip system composed of metal components that minimizes the likelihood for fretting and crevice corrosion?

The inquiry posed above represents a realistic challenge that one might face in the field of orthopedic biomaterials. At a minimum one would want to minimize the galvanic potential of the alloys involved in the implant design. Fretting is the loss of material owing to the rubbing or contact of components; to minimize this type of surface degradation it is necessary to utilize designs that optimize component tolerances for device fixation. Crevice corrosion occurs in metal components that contain fissures, notches, or other geometries that facilitate local changes in the environment. This type of corrosion is further enabled in the presence of elevated stresses or stress concentrations within the component. To minimize fretting and crevice corrosion in a modular hip it is necessary to diminish geometric gaps at junctions such as the Morse Taper (where the stem and acetabular head join) and to keep the stem and neck at reasonable lengths in order to reduce bending stresses at the stem-neck juncture. The case study presented at the end of this chapter addresses this issue.

Historical perspective and overview

Metals have been utilized in medical implants for several hundred years; in fact gold has been used as dental material from as early as 500 BC and noble metals have been employed in structural dentistry for use in fillings, crowns, bridges, and dentures since the 15th century (Williams, 1990). Noble alloys contain varying percentages of gold, palladium, silver, or platinum. High noble metals are metal systems that contain more than 60% noble metals (gold, palladium, or platinum) with at least 40% of the metal being composed of gold. It is known that gold was used in the 16th century to repair cleft palates; gold, bronze, and iron evolved as suture wires over the 17th–18th centuries; and steel plates were commonly employed as internal fixation devices by the early 19th century (Aramany, 1971; Williams et al., 1973). By the mid-1920s, stainless steels were introduced as implant materials and a decade later cobalt-chromium alloys were utilized in the body to provide enhanced corrosion resistance (Park and Bronzino, 2003). Titanium found its use as a surgical material shortly thereafter (Leventhal, 1951). Much of what motivated the development of metals as implant materials was the need for both strength and corrosion resistance in load-bearing devices.

Recommend this book

Email your librarian or administrator to recommend adding this book to your organisation's collection.

Mechanics of Biomaterials
  • Online ISBN: 9780511977923
  • Book DOI: https://doi.org/10.1017/CBO9780511977923
Please enter your name
Please enter a valid email address
Who would you like to send this to *
×
References
Agins, H.J.Alcock, N.W.Bansal, M.Salvati, E.A.Wilson, P.D.Pellicci, P.M.Bullough, P.G. 1988 Metallic wear in failed titanium-alloy total hip replacements. A histological and quantitative analysisJournal of Bone and Joint Surgery 70-A 347
Aramany, M.A. 1971 A history of prosthetic management of cleft palateCleft Palate Journal 8 415
Beddoes, A.Bibby, M.J.Bibby, M.Biddy, M.J. 1999 Principles of Metal Manufacturing ProcessesNew YorkElsevier
Black, J.Skipor, A.Jacobs, J.Urban, R.M.Galante, J.O. 1989 Release of metal ions from titanium-base alloy total hip replacement prosthesesTransactions of the Orthopedics Research Society 14
Collier, J.P.Surprenant, V.A.Jensen, R.E.Mayor, M.B.Surprenant, H.P. 1992 Corrosion between the components of modular femoral hip prosthesesJournal of Bone and Joint Surgery 74B 511
Davies, J.Baldan, N. 1997 Scanning electron microscopy of the bone-bioactive implant interfaceJournal of Biomedical Materials Research 36 429
Fontana, G. 1967 Corrosion EngineeringNew YorkMcGraw Hill
Gilbert, J.L.Buckley, C.A.Jacobs, J.J. 1993 In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion and alloy couplingJournal of Biomedical Materials Research 27 1533
Gilbert, J.L.Jacobs, J.J. 1997 The mechanical and electrochemical processes associated with taper fretting crevice corrosion: a reviewModularity of Orthopedic ImplantsMarlowe, D. E.Parr, J.E.Mayor, M.B.West Conshohocken, PAAmerican Society for Testing and Materials, Special Technical Publication 130145
Hertzberg, R.W. 1989 Deformation and Fracture of Engineering MaterialsNew YorkJohn Wiley & Sons
Hirthe, J.P.Lothe, J. 1982 Theory of DislocationsNew YorkJohn Wiley & Sons
Hull, D.Bacon, D.J. 2001 Introduction to DislocationsLondonButterworth-Heinemann
Hume-Rothery, W. 1960 Atomic Theory for Students of MetallurgyLondonInstitute of Metals
Jacobs, J.J.Skipor, A.K.Doom, P.F.Campbell, P.Schmalzried, T.P.Black, J.Amstutz, H.C. 1996 Cobalt and chromium concentrations in patients with metal on metal total hip replacementsClinical Orthopedics 329 256
Jacobs, J.Gilbert, J.L.Urban, R.M. 1998 Current concepts review – Corrosion of Metal Orthopaedic ImplantsJournal of Bone and Joint Surgery (American) 80 268
Leventhal, G.B. 1951 Titanium, A Metal for SurgeryJournal of Bone and Joint Surgery (American) 33 473
Park, J.B.Bronzino, J.D. 2003 Biomaterials: Principals and ApplicationsNew YorkCRC Press
Pelton, A.R.Duerig, T.Stockel, J. 2004 A guide to shape memory and superelasticity in Nitinol medical devicesMinimally Invasive Therapy and Applied Technology 13 218
Pelton, A.R.Russell, S.M.DiCello, J. 2003 The physical metallurgy of Nitinol for medical applicationsJournal of Minerals, Metals and Materials 55 33
Read, W.T. 1953 Dislocations in CrystalsNew YorkMcGraw-Hill
Richman, M.H. 1974 An Introduction to the Science of MetalsLexington, MAGinn Publishing
Suresh, S. 1998 Fatigue of MaterialsCambridgeCambridge University Press
Weertman, J.Weertman, J.R. 1992 Elementary Dislocation TheoryOxfordOxford University Press
Williams, D.F. 1990 Concise Encyclopedia of Medical and Dental MaterialsOxfordPergamon Press
Williams, D.F.Roaf, R.Maisels, D.O. 1973 Implants in SurgeryLondonW.B. Saunders