Validation of foot-to-foot bioelectrical impedance analysis with dual-energy X-ray absorptiometry in the assessment of body composition in young children: the EarlyBird cohort

  • Joanne Hosking (a1) (a2), Brad S. Metcalf (a1), Alison N. Jeffery (a1), Linda D. Voss (a1) and Terence J. Wilkin (a1)...
Abstract

Foot-to-foot bioelectrical impedance analysis (BIA) is simple and non-invasive, making it particularly suitable for use in children. There is insufficient evidence of the validity of foot-to-foot BIA compared with dual-energy X-ray absorptiometry (DEXA) as the criterion method in healthy young children. Our objective was to assess the validity of foot-to-foot BIA against DEXA in a large cohort of healthy young children. Body composition was measured by foot-to-foot BIA and DEXA in 203 children (mean age 8·9 (sd 0·3) years). Bland–Altman and simple linear regression analyses were used to determine agreement between methods. BIA overestimated fat-free mass by a mean of 2·4 % in boys and 5·7 % in girls, while fat mass was underestimated by 6·5 % in boys and 10·3 % in girls. The percentage fat recorded by BIA was, accordingly, also lower than by DEXA (boys 4·8 %; girls 12·8 %). In boys, however, there were correlations between the size of the difference between methods and the size of the measure under consideration such that in smaller boys fat-free mass was underestimated (r − 0·57; P < 0·001) while fat mass and percentage fat were overestimated (r 0·74 for fat mass; r 0·69 for percentage fat; both P < 0·001) with the reverse in bigger boys. Mean differences between techniques were greater in the girls than in the boys but in boys only, the direction of the differences was dependent upon the size of the child. Therefore, BIA may be useful for large-scale studies but is not interchangeable with DEXA and should be interpreted with caution in individuals.

  • View HTML
    • Send article to Kindle

      To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle.

      Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

      Find out more about the Kindle Personal Document Service.

      Validation of foot-to-foot bioelectrical impedance analysis with dual-energy X-ray absorptiometry in the assessment of body composition in young children: the EarlyBird cohort
      Available formats
      ×
      Send article to Dropbox

      To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your Dropbox account. Find out more about sending content to Dropbox.

      Validation of foot-to-foot bioelectrical impedance analysis with dual-energy X-ray absorptiometry in the assessment of body composition in young children: the EarlyBird cohort
      Available formats
      ×
      Send article to Google Drive

      To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your Google Drive account. Find out more about sending content to Google Drive.

      Validation of foot-to-foot bioelectrical impedance analysis with dual-energy X-ray absorptiometry in the assessment of body composition in young children: the EarlyBird cohort
      Available formats
      ×
Copyright
Corresponding author
* Corresponding author: Dr Joanne Hosking, fax +44 1752 792471, email joanne.hosking@phnt.swest.nhs.uk
References
Hide All
Bland JM & Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i, 307310
Chauhan S, Koo WW, Hammami M & Hockman EM (2003) Fan beam dual energy X-ray absorptiometry body composition measurements in piglets. J Am Coll Nutr 22, 408414
de Onis M & Blossner M (2003) The World Health Organization global database on child growth and malnutrition: methodology and applications. Int J Epidemiol 32, 518526
Ellis KJ (2000) Human body composition: in vivo methods. Physiol Rev 80, 649680
Genton L, Hans D, Kyle UG & Pichard C (2002) Dual-energy X-ray absorptiometry and body composition: differences between devices and comparison with reference methods. Nutrition 18, 6670
Genton L, Karsegard VL, Zawadynski S, Kyle UG, Pichard C, Golay A & Hans DB (2006) Comparison of body weight and composition measured by two different dual energy X-ray absorptiometry devices and three acquisition modes in obese women. Clin Nutr 25, 428437
Goss F, Robertson R, Williams A, Sward K, Abt K, Ladewig M, Timmer J & Dixon C (2003) A comparison of skinfolds and leg-to-leg bioelectrical impedance for the assessment of body composition in children. Dyn Med 2, 5
Heymsfield SB, Lohman TG, Wang Z & Going SB (2005) Human Body Composition. Champaign, IL: Human Kinetics Ltd
Jartti L, Hakanen M, Paakkunainen U, Raittinen P & Ronnemaa T (2000) Comparison of hand-to-leg and leg-to-leg bioelectric impedance devices in the assessment of body adiposity in prepubertal children. The STRIP study. Special Turku coronary Risk factor Intervention Project. Acta Paediatr 89, 781786
Jebb SA, Cole TJ, Doman D, Murgatroyd PR & Prentice AM (2000) Evaluation of the novel Tanita body-fat analyser to measure body composition by comparison with a four-compartment model. Br J Nutr 83, 115122
Kyle UG, Bosaeus I, De Lorenzo AD, et al. (2004a) Bioelectrical impedance analysis – part I: review of principles and methods. Clin Nutr 23, 12261243
Kyle UG, Bosaeus I, De Lorenzo AD, et al. (2004b) Bioelectrical impedance analysis – part II: utilization in clinical practice. Clin Nutr 23, 14301453
Lazzer S, Boirie Y, Meyer M & Vermorel M (2003) Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents. Br J Nutr 90, 987992
Lohman TG & Chen Z (2005) Dual energy X-ray absorptiometry. In Human Body Composition, pp. 6377 [Heymsfield SB, Lohman TG, Wang Z and Going SB, editors]. Champaign, IL: Human Kinetics Ltd
Lukaski HC & Siders WA (2003) Validity and accuracy of regional bioelectrical impedance devices to determine whole-body fatness. Nutrition 19, 851857
Margulies L, Horlick M, Thornton JC, Wang J, Ioannidou E & Heymsfield SB (2005) Reproducibility of pediatric whole body bone and body composition measures by dual-energy X-ray absorptiometry using the GE Lunar Prodigy. J Clin Densitom 8, 298304
Norcross J & Van L (2004) Validation of fan beam dual energy x ray absorptiometry for body composition assessment in adults aged 18–45 years. Br J Sports Med 38, 472476
Nunez C, Gallagher D, Visser M, Pi-Sunyer FX, Wang Z & Heymsfield SB (1997) Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact footpad electrodes. Med Sci Sports Exerc 29, 524531
Parker L, Reilly JJ, Slater C, Wells JC & Pitsiladis Y (2003) Validity of six field and laboratory methods for measurement of body composition in boys. Obes Res 11, 852858
Pietrobelli A, Rubiano F, St-Onge MP & Heymsfield SB (2004) New bioimpedance analysis system: improved phenotyping with whole-body analysis. Eur J Clin Nutr 58, 14791484
Pintauro SJ, Nagy TR, Duthie CM & Goran MI (1996) Cross-calibration of fat and lean measurements by dual-energy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range. Am J Clin Nutr 63, 293298
Ross R & Janssen I (2005) Computed tomography and magnetic resonance imaging. In Human Body Composition, pp. 89108 [Heymsfield SB, Lohman TG, Wang Z and Going SB, editors]. Champaign, IL: Human Kinetics Europe Ltd
Sopher AB, Thornton JC, Wang J, Pierson RN Jr, Heymsfield SB & Horlick M (2004) Measurement of percentage of body fat in 411 children and adolescents: a comparison of dual-energy X-ray absorptiometry with a four-compartment model. Pediatrics 113, 12851290
Sung RY, Lau P, Yu CW, Lam PK & Nelson EA (2001) Measurement of body fat using leg to leg bioimpedance. Arch Dis Child 85, 263267
Tothill P & Hannan WJ (2000) Comparisons between Hologic QDR 1000W, QDR 4500A, and Lunar Expert dual-energy X-ray absorptiometry scanners used for measuring total body bone and soft tissue. Ann N Y Acad Sci 904, 6371
Tyrrell VJ, Richards G, Hofman P, Gillies GF, Robinson E & Cutfield WS (2001) Foot-to-foot bioelectrical impedance analysis: a valuable tool for the measurement of body composition in children. Int J Obes Relat Metab Disord 25, 273278
Voss LD, Kirkby J, Metcalf BS, Jeffery AN, O'Riordan C, Murphy MJ & Wilkin TJ (2003) Preventable factors in childhood that lead to insulin resistance, diabetes mellitus and the metabolic syndrome: the EarlyBird diabetes study 1. J Pediatr Endocrinol Metab 16, 12111224
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

British Journal of Nutrition
  • ISSN: 0007-1145
  • EISSN: 1475-2662
  • URL: /core/journals/british-journal-of-nutrition
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Keywords:

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 25 *
Loading metrics...

Abstract views

Total abstract views: 156 *
Loading metrics...

* Views captured on Cambridge Core between September 2016 - 24th October 2017. This data will be updated every 24 hours.