Hostname: page-component-cf9d5c678-7bjf6 Total loading time: 0.302 Render date: 2021-07-29T14:59:54.895Z Has data issue: true Feature Flags: { "shouldUseShareProductTool": true, "shouldUseHypothesis": true, "isUnsiloEnabled": true, "metricsAbstractViews": false, "figures": true, "newCiteModal": false, "newCitedByModal": true, "newEcommerce": true, "newUsageEvents": true }

Optimization of FIB–SEM Tomography and Reconstruction for Soft, Porous, and Poorly Conducting Materials

Published online by Cambridge University Press:  22 May 2020

Cecilia Fager
Affiliation:
Department of Physics, Chalmers University of Technology, Gothenburg SE-41296, Sweden
Magnus Röding
Affiliation:
RISE Research Institutes of Sweden, Agrifood and Bioscience, Gothenburg, Sweden
Anna Olsson
Affiliation:
AstraZeneca R&D Mölndal, Mölndal SE43183, Sweden
Niklas Lorén
Affiliation:
Department of Physics, Chalmers University of Technology, Gothenburg SE-41296, Sweden RISE Research Institutes of Sweden, Agrifood and Bioscience, Gothenburg, Sweden
Christian von Corswant
Affiliation:
AstraZeneca R&D Mölndal, Mölndal SE43183, Sweden
Aila Särkkä
Affiliation:
Department of Mathematical Sciences, Chalmers University of Technology and University of Gothenburg, Gothenburg, Sweden
Eva Olsson
Affiliation:
Department of Physics, Chalmers University of Technology, Gothenburg SE-41296, Sweden
Corresponding
E-mail address:

Abstract

Tomography using a focused ion beam (FIB) combined with a scanning electron microscope (SEM) is well-established for a wide range of conducting materials. However, performing FIB–SEM tomography on ion- and electron-beam-sensitive materials as well as poorly conducting soft materials remains challenging. Some common challenges include cross-sectioning artifacts, shadowing effects, and charging. Fully dense materials provide a planar cross section, whereas pores also expose subsurface areas of the planar cross-section surface. The image intensity of the subsurface areas gives rise to overlap between the grayscale intensity levels of the solid and pore areas, which complicates image processing and segmentation for three-dimensional (3D) reconstruction. To avoid the introduction of artifacts, the goal is to examine porous and poorly conducting soft materials as close as possible to their original state. This work presents a protocol for the optimization of FIB–SEM tomography parameters for porous and poorly conducting soft materials. The protocol reduces cross-sectioning artifacts, charging, and eliminates shadowing effects. In addition, it handles the subsurface and grayscale intensity overlap problems in image segmentation. The protocol was evaluated on porous polymer films which have both poor conductivity and pores. 3D reconstructions, with automated data segmentation, from three films with different porosities were successfully obtained.

Type
Micrographia
Copyright
Copyright © Microscopy Society of America 2020

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Ballerini, M, Milani, M, Batani, M & Squadrini, F (2001). Focused ion beam techniques for the analysis of biological samples: A revolution in ultramicroscopy? Proc SPIE 4261, 92104.10.1117/12.424523CrossRefGoogle Scholar
Bassim, N, Scott, K & Giannuzi, LA (2014). Recent advances in focused ion beam technology and applications. MRS Bull 39, 317325.10.1557/mrs.2014.52CrossRefGoogle Scholar
Blayvas, I, Bruckstein, A & Kimmel, R (2006). Efficient computation of adaptive threshold surfaces for image binarization. Pattern Recognit 39, 89101.10.1016/j.patcog.2005.08.011CrossRefGoogle Scholar
Breiman, L (2001). Random forests. Mach Learn 45, 532.10.1023/A:1010933404324CrossRefGoogle Scholar
Brostow, W, Gorman, BP & Olea-Mejia, O (2007). Focused ion beam milling and scanning electron microscopy characterization of polymer + metal hybrids. Mater Lett 61, 13331336.CrossRefGoogle Scholar
Cantoni, M & Holzer, L (2014). Advances in 3D focused ion beam tomography. MRS Bull 39(4), 354360.10.1557/mrs.2014.54CrossRefGoogle Scholar
Deng, B, Barki, KG, Ghatak, S, Roy, S, McComb, DW & Sen, CK (2015). FIB/SEM tomography of wound biofilm. Microsc Microanal 21(Suppl 3), 205.10.1017/S1431927615001828CrossRefGoogle Scholar
De Winter, DAM, Schneijdenberg, CTWM, Lebbink, MN, Lich, B, Verkleij, AJ, Druru, MR & Humbel, BM (2009). Tomography of insulating biological and geological materials using focused ion beam (FIB) sectioning and low-kV BSE imaging. J Microsc 233(3), 372383.10.1111/j.1365-2818.2009.03139.xCrossRefGoogle ScholarPubMed
Drobne, D, Milani, M, Leser, V & Tatti, F (2007). Surface damage induced by FIB milling and imaging of biological samples is controllable. Microsc Res Tech 70, 895903.10.1002/jemt.20494CrossRefGoogle ScholarPubMed
Drobne, D, Milani, M, Zrimec, A, Zrimec, MB, Tatti, F & Draslar, K (2005). Focused ion beam/scanning electron microscopy studies of Porcellio scaber (Isopoda, Crustacea) digestive gland epithelium cells. J Scanning Microsc 27, 3034.10.1002/sca.4950270106CrossRefGoogle ScholarPubMed
Dubochet, J, Adrian, M, Chang, J-J, Omo, J-C, Lepault, J, Mcdowall, AW & Schultz, P (1988). Cryo-electron microscopy of vitrified specimens. Q Rev Biophys 21(2), 129228.10.1017/S0033583500004297CrossRefGoogle ScholarPubMed
Efford, N (2000). Digital Image Processing. New York: Addison Wesley.Google Scholar
Egerton, RF (2013). Control of radiation damage in the TEM. Ultramicroscopy 127, 100108.10.1016/j.ultramic.2012.07.006CrossRefGoogle ScholarPubMed
Egerton, RF, Li, P & Malac, M (2004). Radiation damage in the TEM and SEM. Micron 35(6), 399409.CrossRefGoogle ScholarPubMed
Fager, C & Olsson, E (2018). Soft materials and coatings for controlled drug release. In Nanotechnologies in Preventive and Regenerative Medicine, Uskokovic, V & Uskokovic, DP (Eds.), pp. 244259. Netherlands: Elsevier.Google Scholar
Giannuzi, LA & Stevie, FA (2005). Introduction to Focused Ion Beams: Instrumentation, Theory, Techniques and Practise. New York: Springer.10.1007/b101190CrossRefGoogle Scholar
Goldstein, JJ, Newbury, D, Joy, D, Lyman, C, Echlin, P, Lifshin, E, Sawyer, L & Michael, J (2003). Scanning Electron Microscopy and X-ray Microanalysis, 3rd ed. New York: Plenum.CrossRefGoogle Scholar
Hayles, MF, Stokes, DJ, Phifer, D & Findlay, KC (2007). A technique for improved focused ion beam milling of cryo-prepared life science specimens. J Microsc 226(3), 263269.10.1111/j.1365-2818.2007.01775.xCrossRefGoogle ScholarPubMed
Heymann, JA, Hayles, M, Gestmann, I, Giannuzi, LA, Lich, B & Subramaniam, S (2006). Site-specific 3D imaging of cells and tissues with a dual beam, microscope. J Struct Biol 155, 6373.CrossRefGoogle ScholarPubMed
Holzer, L, Indutnyi, F, Gasser, PH, Much, B & Wegmann, M (2004). Three-dimensional analysis of porous BaTiO3 ceramics using FIB nanotomography. J Microsc 216, 8495.CrossRefGoogle ScholarPubMed
Inkson, BJ, Steer, T, Mobus, G & Wagner, T (2001). Subsurface nanoindentation deformation of Cu-Al multilayers mapped in 3D by focused ion beam microscopy. J Microsc 201, 256269.CrossRefGoogle ScholarPubMed
Joergensen, PS, Hansen, KV, Larsen, R & Bowen, JR (2010). A framework for automatic segmentation in three dimensions of microstructural tomography data. Ultramicroscopy 110, 216228.CrossRefGoogle Scholar
Kim, S, Park, MJ, Balsara, NP, Liu, G & Minor, AM (2011). Minimization of focused ion beam damage in nanostructured polymer thin films. Ultramicroscopy 111, 191199.CrossRefGoogle ScholarPubMed
Kirk, ECG, Williamns, DA & Ahmed, H (1987). In situ microsectioning and imaging of semiconductor devices. Inst Phys Conf Ser 87, 691696.Google Scholar
Kizilyaprak, C, Stierhof, YD & Humbel, BM (2019). Volume microscopy in biology: FIB-SEM tomography. Tissue and Cell 57, 123128.10.1016/j.tice.2018.09.006CrossRefGoogle ScholarPubMed
Knapek, E & Dubochet, J (1980). Beam damage to organic material is considerably reduced in cryo-electron microscopy. J Mol Biol 141, 147161.CrossRefGoogle ScholarPubMed
Marko, M, Hsieh, C, Moberlychan, W, Mannella, CA & Frank, J (2006). Focused ion beam milling of vitreous water: Prospects for an alternative to cryo-ultramicrotomy of frozen-hydrated biological samples. J Microsc 222, 4247.10.1111/j.1365-2818.2006.01567.xCrossRefGoogle ScholarPubMed
Marucci, M, Hjärtstam, J, Ragnarsson, G, Iselau, F & Axelsson, A (2009). Coated formulations: New insights into the release mechanism and changes in the film properties with a novel release cell. J Control Release 136(3), 206212.10.1016/j.jconrel.2009.02.017CrossRefGoogle ScholarPubMed
Mayer, J, Giannuzzi, LA, Kamino, T & Michael, J (2007). TEM sample preparation and FIB-induced damage. MRS Bull 32(5), 400407.CrossRefGoogle Scholar
Michael, RJ (2011). Focused ion beam induced microstructural alterations: Texture development, grain growth, and intermetallic formation. Microsc Microanal 17, 386397.CrossRefGoogle ScholarPubMed
Narayan, K & Subramaniam, A (2015). Focused ion beams in biology. Nat Methods 12(11), 10211031.CrossRefGoogle ScholarPubMed
ORS Visual (2018). Object Research Systems [3D reconstruction software]. https://www.theobjects.com.Google Scholar
Parmenter, CDJ, Fay, MW, Hartfield, C & Eltaher, HM (2016). Making the practically impossible “merely difficult” — Cryogenic FIB lift-out for “damage free” soft matter imaging. Microsc Res Tech 79, 298303.10.1002/jemt.22630CrossRefGoogle ScholarPubMed
Salzer, M, Spettl, A, Stenzel, O, Smått, J-H, Lindén, M, Manke, I & Schmidt, V (2012). A two-stage approach to the segmentation of FIB-SEM images of highly porous materials. Mater Charact 69, 115126.CrossRefGoogle Scholar
Schneider, CA, Rasband, WS & Eliceriri, KW (2012). NIH Image to ImageJ: 25 years of image analysis. Nat Methods 9(7), 671675.10.1038/nmeth.2089CrossRefGoogle ScholarPubMed
Schulz, H, Zeile, U, Stodolka, D & Kaft, D (2009). Advantages of a local charge compensation system for FIB/SEM applications on insulating materials. Microsc Microanal 15(Suppl 2), 332.10.1017/S1431927609096639CrossRefGoogle Scholar
Seliger, RL, Ward, JW, Wang, V & Kubena, RL (1979). A high-intensity scanning ion probe with submicrometer spot size. Appl Phys Lett 34(5), 310312.CrossRefGoogle Scholar
Seligman, AM, Wasserkrug, HL & Hanker, JS (1966). A new staining method (OTO) for enhancing contrast of lipid-containing membranes and droplets in osmium tetroxide-fixed tissue with osmiophilic thiocarbohydrazide (TCH). J Cell Biol 30(2), 424432.CrossRefGoogle Scholar
Siepmann, F, Hoffmann, A, Leclercq, B, Carling, B & Siepmann, J (2007). How to adjust desired drug release patterns from ethylcellulose-coated dosage forms. J Controlled Release 119, 182189.CrossRefGoogle ScholarPubMed
Stokes, DJ (2008). Principles and Practice of Variable Pressure: Environmental Scanning Electron Microscopy (VP-ESEM). West Sussex, UK: John Wiley & Sons Ltd.10.1002/9780470758731CrossRefGoogle Scholar
Suzuki, E (2002). High-resolution scanning electron microscopy of immunogold-labelled cells by the use of thin plasma coating of osmium. J Microsc 208, 153157.CrossRefGoogle ScholarPubMed
Tanaka, K & Mitsushima, A (1984). A preparation method for observing intracellular structures by scanning electron microscopy. J Microsc 133, 213222.10.1111/j.1365-2818.1984.tb00487.xCrossRefGoogle ScholarPubMed
Thiedmann, R, Hassfeld, H, Stenzel, O, Koster, LJA, Oosterhout, SD, Van Bavel, SS, Wienk, MM, Loos, J, Janssen, RAJ & Schmidt, V (2011). A multisacle approach to the representation of 3D images, with application to polymer solar cells. Image Anal Stereol 30, 1930.10.5566/ias.v30.p19-30CrossRefGoogle Scholar
Villinger, C, Gregorius, H, Kranz, C, Höhn, K, Münzberg, C, von Wichert, G, Mizaikoff, B, Wanner, G & Walther, P (2012). FIB/SEM tomography with TEM-like resolution for 3D imaging of high-pressure frozen cells. Histochem Cell Biol 138, 549556.10.1007/s00418-012-1020-6CrossRefGoogle ScholarPubMed
Walley, PA, Wineberg, M & Burden, MSTJ (1971). The use of an electron beam evaporation source for electron microscope sample preparation. J Phys E: Sci Instrum 4, 501504.CrossRefGoogle Scholar
Wolff, A, Zhou, Y, Lin, J, Peng, YY, Ramshaw, JAM & Xiao, Y (2018). FIB/SEM processing of biological samples. Microsc Microanal 24(Suppl 1), 822.CrossRefGoogle Scholar
Young, RJ, Dingle, T, Robinson, K & Pugh, PJA (1993). An application of scanned focused ion beam milling to studies on the internal morphology of small arthropods. J Microsc 172, 8188.CrossRefGoogle Scholar
6
Cited by

Send article to Kindle

To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Optimization of FIB–SEM Tomography and Reconstruction for Soft, Porous, and Poorly Conducting Materials
Available formats
×

Send article to Dropbox

To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

Optimization of FIB–SEM Tomography and Reconstruction for Soft, Porous, and Poorly Conducting Materials
Available formats
×

Send article to Google Drive

To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

Optimization of FIB–SEM Tomography and Reconstruction for Soft, Porous, and Poorly Conducting Materials
Available formats
×
×

Reply to: Submit a response

Please enter your response.

Your details

Please enter a valid email address.

Conflicting interests

Do you have any conflicting interests? *