Skip to main content
×
×
Home

In Vivo Biostability of CVD Silicon Oxide and Silicon Nitride Films

  • John M. Maloney (a1), Sara A. Lipka (a1) and Samuel P. Baldwin (a1)
Abstract

Low pressure chemical vapor deposition (LPCVD) and plasma enhanced chemical vapor deposition (PECVD) silicon oxide and silicon nitride films were implanted subcutaneously in a rat model to study in vivo behavior of the films. Silicon chips coated with the films of interest were implanted for up to one year, and film thickness was evaluated by spectrophotometry and sectioning. Dissolution rates were estimated to be 0.33 nm/day for LPCVD silicon nitride, 2.0 nm/day for PECVD silicon nitride, and 3.5 nm/day for PECVD silicon oxide. A similar PECVD silicon oxide dissolution rate was observed on a silicon oxide / silicon nitride / silicon oxide stack that was sectioned by focused ion beam etching. These results provide a biostability reference for designing implantable microfabricated devices that feature exposed ceramic films.

Copyright
References
Hide All
[1] Santini, J. T. Jr , Cima, M. J., and Langer, R., “A controlled release microchip,” Nature, 397, 335338 (1999).
[2] Hämmerle, H., Kobuch, K., Kohler, K., Nisch, W., Sachs, H., and Stelzle, M., “Biostability of micro-photodiode arrays for subretinal implantation,” Biomaterials, 23, 797804 (2002).
[3] Li, Y., Shawgo, R. S., Tyler, B., Henderson, P. T., Vogel, J. S., Rosenberg, A., Storm, P. B., Langer, R., Brem, H., and Cima, M. J., “In vivo release from a drug delivery MEMS device,” J. Controlled Rel., 100, 211219 (2004).
[4] Edell, D. J., “A peripheral nerve information transducer for amputees: long-term multichannel recordings from rabbit peripheral nerves,” IEEE Trans. Biomed. Eng., BME-33, 203214 (1986).
[5] Hetke, J. F., Lund, J. L., Najafi, K., Wise, K. D., and Anderson, D. J., “Silicon ribbon cables for chronically implantable microelectrode arrays,” IEEE Trans. Electron Devices, 41, 314321 (1994).
[6] Osenbach, J. W. and Knolle, W. R., “Behavior of a-SiN-H and a-SiON:H films in condensed water,” J. Electrochem. Soc., 139, 33463351 (1992).
[7] Vogt, M. and Hauptmann, R., “Plasma-deposited passivation layers for moisture and water protection,” Surface Coatings Tech., 74-75, 676681 (1995).
[8] Schmitt, G., Schultze, J.-W., Fassbender, F., Buss, G., Lüth, H., and Schöning, M. J., “Passivation and corrosion of microelectrode arrays,” Electrochimica Acta, 44, 38653883 (1999).
[9] Cogan, S. F., Edell, D. J., Guzelian, A. A., Liu, Y. P., and Edell, R., “Plasma-enhanced chemical vapor deposited silicon carbide as an implantable dielectric coating,” J. Biomed. Mat. Res., 67, 856867 (2003).
[10] Anderson, J. M., “Inflammation, wound healing, and the foreign body response,” in Biomaterials Science, edited by Ratner, B. D. et al, (Academic Press, 1996), pp. 165173.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

MRS Online Proceedings Library (OPL)
  • ISSN: -
  • EISSN: 1946-4274
  • URL: /core/journals/mrs-online-proceedings-library-archive
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed