Utility of Prehospital Quantitative End Tidal CO2?

  • Christopher J. Cooper (a1), James J. Kraatz (a2), David S. Kubiak (a3), James W. Kessel (a1) and Stephen L. Barnes (a1)...
Abstract
Introduction

End tidal CO2 (ETCO2) has been established as a standard for confirmation of an airway, but its role is expanding. In certain settings ETCO2 closely approximates the partial pressure of arterial CO2 (PaCO2) and has been described as a tool to optimize a patient's ventilatory status. ETCO2 monitors are increasingly being used by EMS personnel to guide ventilation in the prehospital setting. Severely traumatized and burn patients represent a unique population to which this practice has not been validated.

Hypothesis

The sole use of ETCO2 to monitor ventilation may lead to avoidable respiratory acidosis.

Methods

A consecutive series of patients with burns or trauma intubated in the prehospital setting over a 24-month period were evaluated. Prehospital arrests were excluded. Absence of ETCO2 transport data and patients without an arterial blood gas (ABG) within 15 minutes of arrival were also excluded. Data collected included demographics, place and time of intubation, service performing intubation, ETCO2 maintained en-route to hospital, and ABG upon arrival. Further data included length of stay, mortality, and injury severity scores.

Results

One hundred sixty patients met the inclusion criteria. Prehospital ETCO2 did not correlate with measured PaCO2 (R2 = 0.08). Mean ETCO2 was significantly lower than mean PaCO2 (34 mmHg vs 44 mmHg, P < .005). Patients arriving acidotic were more likely to die. Mean pH on arrival for survivors and decedents was 7.32 and 7.19 respectively (P < .001). Mortality, acidosis, higher base deficits, and more severe injury patterns were all predictors for a worse correlation between ETCO2 and PaCO2 and increased mean difference between the two values. Decedents and patients presenting with a pH <7.2 demonstrated the greatest discrepancy between ETCO2 and PaCO2. The data suggest that patients may be hypoventilated by prehospital providers in order to obtain a prescribed ETCO2.

Conclusion

ETCO2 is an inadequate tool for predicting PaCO2 or optimizing ventilation in severely injured patients. Adherence to current ETCO2 guidelines in the prehospital setting may contribute to acidosis and increased mortality. Consideration should be given to developing alternate protocols to guide ventilation of the severely injured in the prehospital setting.

CooperCJ, KraatzJJ, KubiakDS, KesselJW, BarnesSL. Utility of Prehospital Quantitative End Tidal CO2?. Prehosp Disaster Med. 2013;28(2):1-6.

Copyright
Corresponding author
Correspondence: Christopher J. Cooper MD Division of Acute Care Surgery University of Missouri One Hospital Drive, MC205 Columbia, MO 65203 USA E-mail coopercj@health.missouri.edu
References
Hide All
1.Caplan, RA, Vistica, MF, Posner, KL, Cheney, FW. Adverse anesthetic outcomes arising from gas delivery equipment: a closed claims analysis. Anesthiology. 1997;87(4):741-748.
2.Bhavani-Shanker, K, Moseley, H, Kumar, AY, Delph, Y. Capnometry and anesthesia. Can J Anaesth. 1992;39:617-632.
3.Cheney, FW, Posner, KL, Lee, LA, Caplan, RA, Domino, KB. Trends in anesthesia related death and brain damage: a closed claims analysis. Anesthesiology. 2006;105:1081-1086.
4.Rozycki, HJ, Sysyn, GD, Marshall, MK, Malloy, R, Wiswell, TE. Mainstream end-tidal carbon dioxide monitoring in the neonatal intensive care unit. Pediatrics. 1998;101:648-653.
5.Morley, TF, Giaimo, J, Maroszan, E, Bermingham, J, Gordon, R, Griesback, R, et al. Use of capnography for assessment of the adequacy of alveolar ventilation during weaning from mechanical ventilation. Am Rev Respir Dis. 1993;148(2):339-344.
6.Ward, KR, Yealy, DM. End-tidal carbon dioxide monitoring in emergency medicine, Part 2: Clinical applications. Acad Emerg Med. 1998;5(6):637-646.
7.Silvestri, S, Ralls, GA, Krauss, B, Thundiyil, J, Rothrock, SG, Senn, A, et al. The effectiveness of out-of-hospital use of continuous end-tidal carbon dioxide monitoring on the rate of unrecognized misplaced intubation within a regional emergency medical services system. Ann Emerg Med. 2005;45(5):497-503.
8.Falk, JL, Rackow, EC, Weil, MH. End-tidal carbon dioxide concentration during cardiopulmonary resuscitation. N Engl J Med. 1988;318(10):607-611.
9.Liu, SY, Lee, TS, Bongard, F. Accuracy of capnography in nonintubated surgical patients. Chest. 1992;102(5):1512-1515.
10.Soto, RG, Fu, ES, Vila, H Jr, Miguel, RV. Capnography accurately detects apnea during monitored anesthesia care. Anesth Analg. 2004;99(2):379-382.
11.Krauss, B, Hess, DR. Capnography for procedural sedation and analgesia in the emergency department. Ann Emerg Med. 2007;50(2):172-181.
12.Fuke, S, Miyamoto, K, Ohira, H, Ohira, M, Odajima, N, Nishimura, M. Evaluation of transcutaneous CO2 responses following acute changes in PaCO2 in healthy subjects. Respirology. 2008;14:436-442.
13.Wimberley, PD, Pedersen, KG, Thode, J, Fogh-Anderson, N, Sorensen, AM. Transcutaneous and capillary PCO2 and PO2 measurements in healthy adults. Clin Chem. 1983;29:1471-1473.
14.Sanders, MH, Kern, NB, Costantino, JP, Stiller, RA, Strollo, PJ. Accuracy of end-tidal and transcutaneous PCO2 monitoring during sleep. Chest. 1994;106:472-483.
15.Warner, KJ, Cuschieri, J, Garland, B, Carlbom, D, Baker, D, Copass, MK, et al. The utility of end-tidal capnography in monitoring ventilation status after severe injury. J Trauma. 2009;66:26-31.
16.Chesnut, RM, Marshall, LF, Klauber, MR. The role of secondary brain injury in determining outcome from severe head injury. J Trauma. 1993;34:216-222.
17.Warner, KJ, Cuscheri, J, Copass, MK, Jurkovich, GJ, Bulger, EM. Emergency department ventilation effects outcome in severe traumatic brain injury. J Trauma. 2008;64:341-347.
18.Davis, DP, Idris, AH, Sise, MJ, et al. Early ventilation and outcomes in patients with moderate to severe traumatic brain injury. Crit Care Med. 2006;34:1202-1208.
19.Salomone, JP, Pons, PT, McSwain, NE, et al, eds. Prehospital Trauma Life Support, 6th ed. St. Louis, Missouri USA: Mosby; 2006.
20.Shankar, KB, Moseley, H, Vemula, V, Ramasamy, M, Kumar, Y. Arterial to end-tidal carbon dioxide tension difference during anesthesia in early pregnancy. Can J Anaesth. 1989;36:124-127.
21.Fletcher, R, Johnson, B. Deadspace and the single breath test for carbon dioxide during anaesthesia and artificial ventilation. Effects of tidal volume and frequency of respiration. Br J Anaesth. 1984;56:109-119.
22.Hopper, AO, Nystrom, GA, Deming, DD, Brown, WR, Peabody, JL. Infrared end-tidal CO2 measurement does not accurately predict arterial CO2 values or end-tidal to arterial PCO2 gradients in rabbits with lung injury. Pediatr Pulmonol. 1994;17(3):189-196.
23.Prause, G, Hetz, H, Lauda, P, Pojer, H, Smolle-Juettner, F, Smolle, J. A comparison of the end-tidal-CO2 documented by capnometry and the arterial pCO2 in emergency patients. Resuscitation. 1997;35:145-148.
24.Belpomme, V, Ricard-Hibon, A, Devoir, C. Correlation of arterial PCO2 and PETCO2 in prehospital controlled ventilation. Am J Emerg Med. 2005;23:852-859.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Prehospital and Disaster Medicine
  • ISSN: 1049-023X
  • EISSN: 1945-1938
  • URL: /core/journals/prehospital-and-disaster-medicine
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Keywords

Metrics

Altmetric attention score

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed