We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
Online ordering will be unavailable from 17:00 GMT on Friday, April 25 until 17:00 GMT on Sunday, April 27 due to maintenance. We apologise for the inconvenience.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
To determine whether reprocessed single-use devices would meet regulatory standards for sterility and meet the same materials standards as a new device.
Design:
The study included single-use and reusable biopsy forceps and papillotomes and a reusable stone retrieval basket. The suitability of these devices for cleaning and disinfection or sterilization was examined.
Methods:
Testing of cleanability was conducted on devices contaminated with technetium 99-radiolabeled human blood. Instruments were cleaned using hospital recommended practices for manual cleaning. Gamma counts per second were determined before and after cleaning to localize contaminants, which were additionally visualized using light and scanning electron microscopy. X-ray phc-toelectron spectroscopy was used to quantify contamination elements on the materials tested. Residual bioburden testing on instruments contaminated with microorganisms suspended in coagulable sheep blood was carried out to establish the efficacy of disinfection and sterilization.
Results:
All devices remained contaminated after cleaning, but single-use devices and the stone basket tended to be more heavily contaminated than reusable forceps and papillotomes. Cleaning procedures facilitated distribution of contaminants further into the lumens of the disposable forceps. Decreased concentrations of silicon and increased concentrations of carbon and nitrogen suggested that layers of silicon lubricant had been removed and contaminants were organic material. Reusable devices were effectively disinfected, but single-use devices were not Sterilization could not eliminate the challenge microorganisms completely.
Conclusions:
None of the reprocessed single-use instruments were effectively cleaned, disinfected, or sterilized. This condition may provide an opportunity for the viability of non-resistant or nosocomial organisms and viruses. Additionally, reprocessing procedures may result in material destruction of fragile devices. Cost-saving initiatives that have inspired reprocessing of single-use devices, despite the absence of data establishing the efficacy of decontamination and the durability of materials throughout reprocessing, should be pushed into the background.