To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Transcatheter closure of large and complex atrial septal defect can pose challenges and complications during device placement. To improve stability, several assistive techniques have been developed.
Methods:
This retrospective study evaluated the efficacy of the device-assisted device closure technique for large secundum atrial septal defects. Patients who underwent device-assisted device closure of atrial septal defect between December 2023 and August 2024 were analysed.
Results:
Twenty patients (mean age 38.69 years) underwent device closure of large secundum atrial septal defect with device-assisted device closure technique. The mean atrial septal defect diameter was 31.9 mm. The average thick-to-thick measurement was 38.3 mm, which determined the device size. The majority (18 cases) had thin, floppy margins and two had deficient inferior rim. Successful closure was achieved in 18 patients (90%), while two patients (10%) required other methods of assistance. Based on fluoroscopic guidance, patients were divided into two groups: Group A (8 patients) used anteroposterior projection, and Group B (12 patients) used left anterior oblique-cranial view. After initial two failures with anteroposterior view, all cases were successfully closed using left anterior oblique-cranial projection. Device sizes ranged from 36 to 50 mm (median 40 mm). Cocoon devices were used for sizes up to 42 mm, and Occlunix for larger devices. No significant procedural complications occurred, although two patients had minor post-procedural events.
Conclusions:
Device-assisted device closure technique offers a promising and safe dynamic assistance approach for transcatheter closure of large and challenging atrial septal defects. The left anterior oblique-cranial view showed promising results, though without statistical significance. While results are encouraging, larger prospective studies are needed to validate its effectiveness.
Lifetech Konar-multifunctional occluder is a novel device which is primarily used for the closure of ventricular septal defects. Being “multifunctional”, the occluder has the potential to be useful in various structural cardiac defects.
Materials and methods:
We share our retrospective review from two centres regarding non-conventional usage of multifunctional occluders in CHD. Eight patients who underwent interventions using multifunctional occluders for lesions other than ventricular septal defects between March 2019 to September 2019 were included in the study. The patients were analysed based on demography, the size and type of lesion, procedural success, and development of complications. All patients were followed up in the outpatient department for a minimum period of 6 months.
Results:
The median age and weight of the cohort were 3.2 years and 9 kg, respectively. Six patients had patent ductus arteriosus, while one patient had aorto-pulmonary window and one had a coronary arterio-venous fistula. The sizing of the occluders and the procedural approach were based on the underlying pathology. The most commonly used occluder was 6 × 4 mm variant. One patient had successful implantation but had significant intra-device residual flow and was thus replaced by a different occluder. There were no major complications, nor any incidences of device embolisation or malposition. On follow-up, all patients had uneventful course.
Conclusion:
Konar-multifunctional occluder can be safely used in lesions other than ventricular septal defects, when needed under specific circumstances. Its unique characteristics make it a versatile choice in a variety of cardiac lesions.