Hostname: page-component-7c8c6479df-5xszh Total loading time: 0 Render date: 2024-03-28T08:42:12.546Z Has data issue: false hasContentIssue false

Experimental study on heat transmission to the vestibule during CO2 laser use in revision stapes surgery

Published online by Cambridge University Press:  01 November 2006

M Szymański
Affiliation:
Department of Otolaryngology, Head and Neck Surgery, Medical University of Lublin, Poland
K Morshed
Affiliation:
Department of Otolaryngology, Head and Neck Surgery, Medical University of Lublin, Poland
R Mills
Affiliation:
Otolaryngology Unit, University of Edinburgh, Edinburgh, Scotland, UK

Abstract

We studied the transmission of heat to the vestibule during revision stapes surgery with a piston in situ, using a CO2 laser, in an in vitro model. A type K thermocouple was placed around the medial end of stainless steel and fluoroplastic wire pistons in a ‘vestibule’ filled with saline. The effect of laser hits on fluoroplastic wire and stainless steel stapes prostheses was investigated. The effect of introducing a vein graft to seal the stapedotomy was also examined. Greater temperature rises occurred with stainless steel than with fluoroplastic wire pistons. The addition of the vein graft reduced heat transmission.

Application of the CO2 laser to fluoroplastic wire pistons, using the power settings suggested by the manufacturer, is not likely to damage the inner-ear structures. Application of 6 W laser energy to stainless steel pistons can potentially disturb the inner-ear function.

Type
Main Articles
Copyright
2007 JLO (1984) Limited

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)