A new continuous high-resolution detection system for sulphate in ice cores

Published online by Cambridge University Press:  14 September 2017

Matthias Bigler
Affiliation:
Ice and Climate Research, Niels Bohr Institute, University of Copenhagen, Juliane Maries Vej 30, DK-2100 København Ø, Denmark E-mail: bigler@gfy.ku.dk
Anders Svensson
Affiliation:
Ice and Climate Research, Niels Bohr Institute, University of Copenhagen, Juliane Maries Vej 30, DK-2100 København Ø, Denmark E-mail: bigler@gfy.ku.dk
Jørgen Peder Steffensen
Affiliation:
Ice and Climate Research, Niels Bohr Institute, University of Copenhagen, Juliane Maries Vej 30, DK-2100 København Ø, Denmark E-mail: bigler@gfy.ku.dk
Patrik Kaufmann
Affiliation:
Climate and Environmental Physics, Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Rights & Permissions [Opens in a new window]

Abstract

Sulphate (SO42–) is a major ion found in polar ice cores and is related to different aerosol sources and processes. Explosive volcanic eruptions, even far away, can cause distinct sulphate peaks in ice core records. Thus, a robust sulphate detection system which is suitable for fieldwork and which enables the measurement of sulphate at high temporal resolution is of great interest. In this study, we present the adaptation of a new continuous flow analysis system for sulphate that is based on a spectrophotometric method using dimethylsulfonazo III and an inline reactor column containing barium sulphate particles. The method shows a detection limit of ∽70 ng g–1 and a linear range up to at least 3000 ng g–1. It is simple, robust and less prone to interferences compared to the previously used method.

Type
Research Article
Copyright
Copyright © The Author(s) [year] 2007

Introduction

Ice core records of irreversibly deposited aerosol constituents carry comprehensive information on changes in their atmospheric long-range transport and source areas far back in time (Reference Legrand and MayewskiLegrand and Mayewski, 1997; Reference WolffWolff and others, 2006). Measured at high temporal resolution they may reveal even persistent seasonal deposition cycles (Reference Sommer, Wagenbach, Mulvaney and FischerSommer and others, 2000; Reference BiglerBigler and others, 2002), a prerequisite for stratigraphic dating (Reference RasmussenRasmussen and others, 2006). Some of the ice core records have distinct peaks related to aerosols superimposed, caused by explosive volcanic eruptions. Thus, such time series can be used to estimate occurrence rates of volcanic events to assess possible climatic effects (Reference CrowleyCrowley, 2000; Reference Hegerl, Crowley, Hyde and FrameHegerl and others, 2006). Furthermore, these unique archives of past volcanic activity can serve to pin down the dating of ice cores and other paleoarchives by identifying common reference horizons. Explosive volcanic eruptions emit large amounts of ash particles (e.g. tephra) and gases (mainly SO2) into the atmosphere, where oxidation and gas-to-particle conversion of SO2 to sulphuric acid (H2SO4) occurs. Volcanic events are therefore recorded in ice cores mainly as acidic peaks (Reference Hammer, Clausen and DansgaardHammer and others, 1980; Reference ClausenClausen and others, 1997), sulphate (SO4 2–) peaks (Reference Zielinski, Mayewski, Meeker, Whitlow and TwicklerZielinski and others, 1996; Reference BiglerBigler and others, 2002; Castelleno and others, 2005) and tephra (Reference GrönvoldGrönvold and others, 1995; Reference Mortensen, Bigler, Grönvold, Steffensen and JohnsenMortensen and others, 2005). Continuous high-resolution measurements are not feasible for tephra, but they can be carried out for SO4 2– and the acidic content of the ice. However, the latter is only indirectly possible through conductivity measurements on solid ice (Electrical Conductivity Measurements (ECM) and Dielectric Profiling (DEP)) or on meltwater of continuously melted ice samples (electrolytical conductivity). Both ECM and DEP can be weakened or even obliterated in sections with high dust levels due to the increased alkalinity, whereas electrolytical conductivity can result in ambiguous peaks, as it contains a bulk signal of all ionic components present in the ice. Therefore, large SO4 2– peaks must be considered as the most reliable indicator of volcanic activity recorded in ice cores.

Ion chromatography (IC) is a common and very reliable method to measure SO4 2– in ice core samples (Reference Legrand, Delmas, Jones and Orville-ThomasLegrand and Delmas, 1987; Reference LittotLittot and others, 2002). However, it is time consuming to measure numerous samples to achieve high-resolution records, especially when each sample has to be cut and decontaminated manually. Therefore, the Continuous Flow Analysis (CFA) measuring technique has become the method of choice to obtain long ice core time series of several important aerosol constituents with a depth resolution of ∽1 cm, which leads to an unprecedented temporal resolution (considering the actual annual layer thickness) (Reference Sigg, Fuhrer, Anklin, Staffelbach and ZurmuehleSigg and others, 1994; Reference RöthlisbergerRöthlisberger and others, 2000; Reference Bigler, Röthlisberger, Lambert, Stocker and WagenbachBigler and others, 2006). The strength of a CFA system is twofold: (1) The melting device provides a continuous uncontaminated sample stream of meltwater from continuously melted subsections of the ice core. This is because only the inner part of the subsection that was never in contact with ambient air is drained off, whereas the possibly contaminated water from the outer part is rejected. (2) The immediate detection of the constituents of interest by means of fluorescence and absorption spectrophotometric methods leads to a very high temporal data resolution. Although the CFA melting device was recently used for IC measurements, either by using an autosampler or by direct coupling to an adapted fast IC (Reference UdistiUdisti and others, 2000; Reference Osterberg, Handley, Sneed, Mayewski and KreutzOsterberg and others, 2006), the CFA detection methods are still unprecedented in terms of temporal data resolution with respect to the length of ice core records. Therefore the availability of a CFA SO4 2– detection system is highly desirable. A first attempt at measuring SO4 2– by CFA was made some years ago (Reference BiglerBigler, 2000; Reference BiglerBigler and others, 2002) and records were obtained from the glacial (∽110 to ∽10 kyr BP) part of the North Greenland Ice Core Project (NGRIP) deep ice core (Reference BiglerBigler, 2004; North Greenland Ice Core Project members, 2004) and from a shallow firn core taken in Northeast Greenland (NGT B20) (Reference BiglerBigler and others, 2002). However, the method used for the SO4 2– detection based on Methylthymol Blue (MTB) (Reference Madsen and MurphyMadsen and Murphy, 1981) turned out to be relatively delicate for routine field measurements, especially at low concentration levels, showing occasional baseline instabilities, a non-linear calibration curve, a too high Limit of Detection (LOD) and strong interferences from bivalent cations such as calcium and magnesium. Ideally, the new CFA SO4 2– detection system should cover a concentration range of SO4 2– in polar ice cores from 10 to 10 000 ng g–1. Here we present and discuss the adaptation of a new CFA SO4 2– detection method.

Method and Optimization

Overview

Based on an extensive literature investigation, a Flow Injection Analysis (FIA) method using dimethylsulfonazo III (DMS-III) as SO4 2– indicator (Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004) was found to be most promising as a starting point for our adaptation to a CFA method. A similar FIA method was used already by Reference Yang, Zhang and KorenagaYang and others (1997a, Reference Yang, Zhang, Korenaga and Higuchib), with reference to the initial publication of Reference Kondo, Miyata and TôeiKondo and others (1982) using DMS-III. The FIA DMS-III method has many important features. Depending on the SO4 2– concentration in the sample, barium sulphate (BaSO4) precipitates and causes the dissociation of Ba-DMS-III complexes. The decreasing absorbance at 662 nm is monitored continuously. An inline reactor column containing BaSO4 particles (Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004) promotes the efficiency and reproducibility of the slow liquid-solid reaction and the radial mixing of sample and reagent (Reference Yang, Zhang and KorenagaYang and others, 1997a). Furthermore, the sample water has to be pre-treated in an inline Cation Exchange Column (CEC) to avoid interferences with specific bivalent cations (i.e. Ca2+ and Mg2+). The given LOD ranges from 30 ng g–1 (Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004) to 80 ng g–1 (Reference Yang, Zhang and KorenagaYang and others, 1997a), while the precision (relative standard deviation of repeated similar measurements) is found to be between 1% (Reference Kondo, Miyata and TôeiKondo and others, 1982) and 1.6% (Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004) for concentrations of ∽1000 ng g–1 SO42–. Figure 1 provides an overview of the new CFA SO4 2– set-up and each of the components are described in the following sections.

Fig. 1. Setup of the CFA SO4 2– detection system. Features include the inline Cation Exchange Column (CEC), the overflow debubbler and the inline BaSO4 reactor column. Flow rates are given for each pump tube of the peristaltic pump. The broken line indicates the fibre optic cables from the flow cell to the lamp and to the detector, respectively.

Reagent

Barium (Ba2+) reacts with DMS-III at an equimolar ratio. According to Reference Kondo, Miyata and TôeiKondo and others (1982) it is appropriate to fix the molar ratio of Ba2+ to DMS-III in the reagent at 0.7–1. The high ethanol content of the reagent leads to an increased absorbance, while the pH value of the merged sample and reagents flow has to be kept between 3.5 and 5.5 with an equimolar solution of chloroacetic acid (ClCH2COOH) and sodium chloroacetate (ClCH2COONa), which is a buffer with pH 2.9. The addition of potassium nitrate (KNO3) to the reagent maintains a constant ionic strength, which prevents baseline fluctuations. However, the Ba-DMS-III concentration in the reagent is the most important parameter and can be used to optimize the method for the desired SO4 2– concentration range. While low reagent concentrations cause a low signal-to-noise ratio, too high reagent concentrations lead to deviations of the calibration curve from linearity as well as a lower sensitivity for low SO4 2– concentration levels and therefore a higher LOD. We found suitable concentrations of 80 mM for DMS-III and 56 mM for BaCl2 in the reagent somewhat higher than in the initial publications (Reference Kondo, Miyata and TôeiKondo and others, 1982; Reference Yang, Zhang and KorenagaYang and others, 1997a; Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004). The reagent preparation is described in the following paragraph.

All reagents were of analytical grade and prepared with ultra pure water (having a conductivity of 18MΩcm–1). Included are: 8mL 0.01 M DMS-III solution (M = 760.66 g mol–1, Sigma-Aldrich); 5.6 mL 0.01 M BaCl2 solution (M = 208.25 g mol–1, Merck); 5 mL 1 M KNO3 solution (M = 101.11 g mol–1, Merck); 20 mL buffer solution (pH 2.9) consisting of a 1:1 mixture (K. Higuchi, personal communication) of 1M chloroacetic acid (M = 94.50 g mol–1, Merck) and 1 M sodium chloroacetate (M = 116.48 g mol–1, Merck); 700 mL ethanol absolute (Merck); and ∽260mL ultra pure water to get 1000mL reagent solution. It is relatively stable when ultraviolet light is avoided (S. Motomizu, personal communication, 2005). A SO4 2– stock standard solution of Na2SO4 in water (Merck) was used to dilute appropriate working standard solutions.

BaSO4 reactor column

Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others (2004) packed crushed BaSO4 with a particle size of about 90 mm in their reactor column. However, this material was neither commercially available from fine chemical suppliers, nor could we get detailed information about possible extraction procedures from the authors. BaSO4 is a white powder of low solubility and high density and is used as a non-fading and chemically stable pigment, roentgen contrast medium and filler material in paper, rubber or heavy concrete. It is produced by disintegration of the mineral barite (Schwerspat) and controlled precipitation of Ba2+ salts with sulphuric acid. A method of producing particle sizes between 0.1 and 10 mm is given by Reference Amirzadeh-Asl and SchwarzAmirzadeh-Asl and Schwarz (2001). Because of the larger size, we therefore assume that the above-mentioned crushed BaSO4 has to be barite mineral particles and not precipitated BaSO4. A barite test sample was kindly provided by Sachtleben Chemie (Germany), mined from the Clara Mine deposit at Wolfach. The crude ore was dressed by heavy media separation, flotation and drying processes. The specific particle distribution, with a medium particle size of ∽23 mm, was achieved by milling and air sifting. The final product has a BaSO4 content of ∽98% (K.-H. Schwarz, personal communication, 2006). By sieving with a 90 mm sieve we obtained the fraction of larger particles (∽30% of the total amount) and washed it 10 times with ultra pure water. Subsequently, reactor columns with 1.6 mm Inner Diameter (ID) were filled (details below). We tested different column lengths of 4– 24cm and found that 15cm is suitable for our set-up. Shorter columns seem to enhance baseline drifts, while longer ones can cause pressure problems. We observed that the length of the column does not affect the absorbance very much. As slight baseline drifts are observed after ∽10– 12 hours of usage, the column has to be rinsed with ultra pure water for 1 hour. Subsequently, the baseline returns to the initial level again. A column can be used for 1 month with unchanged sensitivity and reproducibility (Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004).

Cation exchange column (CEC)

In contrast to monovalent ions, bivalent cations such as calcium (Ca2+) and magnesium (Mg2+) interfere with the detection method (Reference Kondo, Miyata and TôeiKondo and others, 1982; Reference Yang, Zhang and KorenagaYang and others, 1997a; Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others, 2004). Therefore, an inline CEC has to be used. Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others (2004) reported that Na+-type cation exchange resins yield better results than those of H+-type. As a consequence, we used Dowex 50W×8, Na+-form with a particle size of 75–150 mm and analytical grade (Fluka) in a 15 cm long, 1.6 mm ID column (details below). The CEC can be regenerated by pumping through a 20% sodium chloride solution for a few minutes, which was performed daily. The CEC was tested and found to be appropriate to prevent interferences caused by bivalent cations in typical glacial ice core samples from Greenland. A solution containing 2000 ng g–1 Ca2+ and 1000 ng g–1 Mg2+, never observed in high-resolution CFA data from polar ice cores previously (Reference BiglerBigler, 2004), caused no noticeable interference. A smaller column volume could be chosen for interglacial or Antarctic ice samples with lower impurity content.

Set-up

A characteristic feature of the applied SO4 2– detection method is the formation of bubbles due to heat production when merging the ethanolic reagent with the sample water stream. During the experimental testing it became apparent that bubble formation has to be prevented by all means as it causes a very noisy signal and baseline instabilities. This problem was solved by (1) using a lower ethanol concentration in the reagent compared to the initial publication (Reference Kondo, Miyata and TôeiKondo and others, 1982), as proposed by Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others (2004); (2) inserting an overflow debubbler into the set-up, which removes large bubbles and (3) by adding a backpressure tube after the flow cell, which prevents additional small bubbles from forming in the light path of the flow cell. According to the law of Hagen-Poiseuille the difference pressure of a 25 cm long backpressure tube with an ID of 0.25 mm is ∽0.5 bar at a flow rate of 1.1 mLmin–1 (estimated dynamic viscosity of an ethanol-water mixture at 208C is 0.0011 Nsm–2). Altogether, the two inline columns and the backpressure tube in the set-up cause a high backpressure, so that the peristaltic pumps run at their pressure limit and exhibit pulsating pumping. This leads to irregular mixing of sample and reagents and therefore noisy signals. This was a further reason to decouple the set-up by inserting an overflow debubbler. Additionally, a mixing coil with an optimized length of 100 cm in front of the debubbler was found to support a more regular mixing.

The complete set-up shown in Figure 1 was made of Teflon tubing (1.6mm Outer Diameter (OD), 0.5 mm ID, PFA HP plus). Due to the relatively high pressure within the whole set-up, all connections were screwed using standard 1/4-28 fittings and peristaltic tubing adapters (all items from Upchurch, USA). Flow rates (given in Fig. 1) were adjusted using pump tubes with different inner diameters and appropriate pump rates of the peristaltic pumps (from Ole Dich, Denmark). The pump tubes are solvent-resistant for ethanol containing streams (PharMed/Ismapren) and made of Tygon Long Flex Life (LFL) for all other lines (both from Ismatec, Switzerland). They were checked regularly. We found that changing the flow rate did not have a remarkable effect, except for reducing absorbance values by a small amount at lower flow rates. The overflow debubbler was made of a downwards-mounted pipette tip with the sample inlet and overflow tubes inserted from the top and the outlet connected to the tip at the bottom. The overflow tube was positioned slightly higher than the sample inlet tube to prevent contamination. By shifting the position of these two tubes it was possible to reduce the mixing volume to 40 mL while still maintaining an efficient removal of bubbles in the flow. Columns were built by using standard 1/4-28 fittings and unions to connect the flow lines with the column tubes made of Teflon tubing (3.2mm OD, 1.6 mm ID, PFA HP plus). On both sides a frit-in-a-ferrule (for 3.2 mm OD, 2 mm SST, PCTFE) was used to keep the filling in the columns (all items from Upchurch, USA). To run standard series automatically and to switch between sample and blank, actuated medium pressure ceramic valves (V-1461-DC 6-port-2-position and V-1471-DC 7-port-6-position from Upchurch, USA) were used.

Absorption detector

A miniature combination deuterium and tungsten halogen light source (DT-Mini-2-GS from FIAlab/Ocean Optics, USA) was employed. However, only the halogen bulb was switched on for this application. To detect the absorption signal a flow cell with a 10mm optical path (SMA-Z-mvol, Ultem, with fused silica windows, from FIAlab, USA) was used. Fibre optic cables conducted the light from the lamp to the flow cell and from there to the spectrometer. The detector consisted of one channel of a three-channel spectrometer (USB2000-UV-VIS, ADC1000-USB analogue-digital converter from FIAlab/Ocean Optics, USA). We made use of the automatic correction function for electrical dark signals. Maximum absorbance was found at a wavelength of 659 nm (insertion in Fig. 2), slightly lower than in the initial publications. However a maximum signal-to-noise ratio was achieved by integrating the spectrometer signal over the wavelength interval 654–663 nm. All connections to the measuring computer were made directly or via RS-233 to USB converters to one USB hub. The control and data acquisition program with a measuring sequence time of 1 s was written in LabVIEW 8 (National Instruments, USA) using the driver software OOILVD (from Ocean Optics/FIAlab) to control the spectrometer.

Fig. 2. (a) Spectrometer signal S (number of counts) of a SO4 2– standard series (dark grey) with given concentration values. The baseline (light grey) is calculated by linearly interpolating between two sections of blank measurements (black) before and after the standards. (b) Calculated absorption values A of the standard series (dark grey) with respect to the zero line (light grey). Plateau values used to calculate the calibration curve (Fig. 3) are highlighted in black. The insertion shows the measured absorption spectrum A (in arbitrary units) with a maximum around the wavelength λ = 659 nm.

Characteristics of the new method

Figure 2 shows the spectrometer signal S of a SO4 2– standard series. The baseline S 0 is calculated by linearly interpolating between two sections of blank measurements (ultra pure water) at the beginning and the end. The absorption A (Fig. 2b) is calculated as A = log10 (S 0 /S) = ∊lC, where is the extinction coefficient in (ng g–1)–1cm–1, l the light path length in the flow cell in cm and C the SO4 2– concentration in ng g–1. Plateau values of the standards define the calibration curve shown in Figure 3. The characteristics of our new SO4 2– method are an extinction coefficient of ∽0.0001 (ng g–1)–1 cm–1 and a LOD of ∽70 ng g–1, approximately double the value of 30 ng g–1 given in Reference Burakham, Higuchi, Oshima, Grudpan and MotomizuBurakham and others (2004) and higher than desired for our CFA SO4 2– detection method to measure polar ice samples. However, a value of this order of magnitude is probably the lowest possible due to the solubility of BaSO4 in aqueous solution (S. Motomizu, personal communication, 2005). The upper limit of the linear calibration curve is ∽3000 ng g–1, somewhat lower than desired. Nevertheless, it still covers almost all the volcanic peak values expected in polar ice cores. By applying a non-linear calibration curve or by using a higher Ba-DMS-III concentration in the reagent (with the disadvantage of a lower sensitivity at low concentration levels) this upper limit could be raised.

Fig. 3. Calibration curve between measured absorption values and given standard concentrations (points). To calculate the linear fit (A = –0.000 103 C) the two highest values were excluded to account for the deviation from linearity at high concentration levels. The insertion contains an enlargement of the lowest concentration values showing a satisfactory linearity (error bars correspond to the standard deviations of the standard plateaus in Fig. 2).

Conclusions

A new CFA SO4 2– detection method has been successfully implemented. It is based on an existing FIA method using Ba-DMS-III complexes in the spectrophotometric reagent. The handling of this new method is simpler than that of the previously used CFA SO4 2– method, especially concerning the reagent preparation. Furthermore, the reagent composition can be adjusted so that a linear calibration curve, also for low concentration levels, is achieved, leading to more accurate and simpler data processing. Compared to the previous method, we found a more stable baseline and a greater suppression of interferences with the volume of the cation exchange column used here. This leads to a more robust application of the method and to more accurate results.

However, the detection limit of the new method is in the same order of magnitude as for the previous method, which is still too high for low background concentration levels. Therefore our method is sufficient and suitable for peak monitoring, but not good enough for accurate background measurements on polar ice samples, especially from interglacial periods.

Despite this drawback, further investigations using the new method will be performed due to the lack of a more sensitive method. To assess the quality of the new method, selected sections of the NGRIP ice core will be re-measured and the new results compared to the values obtained with the previously used CFA SO4 2– method. We will also try to perform corrections for the asymmetric shape of standard signals and peaks by applying mathematical reso lution enhancement techniques such as back-diffusion (Reference Rasmussen, Andersen, Johnsen, Bigler and McCormackRasmussen and others, 2005) or Monte Carlo approaches.

Acknowledgements

This work is part of the Copenhagen Ice Core Dating Initiative which is supported by a grant from the Carlsberg Foundation. We would like to thank K.-H. Schwarz from Sachtleben Chemie GmbH, Germany, for providing a barite test sample.

References

Amirzadeh-Asl, D. and Schwarz, K.-H.. 2001. Sachtleben Chemie. DE Pat. 10005685. Germany.Google Scholar
Bigler, M. 2000. Entwicklung und Anwendung einer neuen Methode zur kontinuierlichen, hochaufgelösten Messung der Sulfatkonzentration an alpinen und polaren Eisbohrkernen. (MSc thesis, University of Bern.)Google Scholar
Bigler, M. 2004. Hochauflösende Spurenstoffmessungen an polaren Eisbohrkernen: Glazio-chemische und klimatische Prozessstudien. (PhD thesis, University of Bern.)Google Scholar
Bigler, M. and 6 others. 2002. Sulphate record from a northeast Greenland ice core over the last 1200 years based on continuous flow analysis. Ann. Glaciol., 35, 250–256.Google Scholar
Bigler, M., Röthlisberger, R., Lambert, F., Stocker, T.F. and Wagenbach, D.. 2006. Aerosol deposited in East Antarctica over the last glacial cycle: detailed apportionment of continental and sea-salt contributions. J. Geophys. Res., 111, D08205. (10.1029/ 2005JD006469.)Google Scholar
Burakham, R., Higuchi, K., Oshima, M., Grudpan, K. and Motomizu, S.. 2004. Flow injection spectrophotometry coupled with a crushed barium sulfate reactor column for the determination of sulfate ion in water samples. Talanta, 64(5), 1147–1150.Google Scholar
Castellano, E. and 9 others. 2005. Holocene volcanic history as recorded in the sulfate stratigraphy of the European Project for Ice Coring in Antarctica Dome C (EDC96) ice core. J. Geophys. Res., 110, D06114. (10.1029/2004JD005259.)Google Scholar
Clausen, H.B. and 6 others. 1997. A comparison of the volcanic records over the past 4000 years from the Greenland Ice Core Project and Dye 3 Greenland ice cores. J. Geophys. Res., 102(C12), 26,707–26,723.Google Scholar
Crowley, T.J. 2000. Causes of climate change over the past 1000 years. Science, 289(5477), 270–277.Google Scholar
Grönvold, K. and 6 others. 1995. Ash layers from Iceland in the Greenland GRIP ice core correlated with oceanic and land sediments. Earth Planet. Sci. Lett., 135(1–4), 149–155.CrossRefGoogle Scholar
Hammer, C.U., Clausen, H.B. and Dansgaard, W.. 1980. Greenland ice sheet evidence of post-glacial volcanism and its climatic impact. Nature, 288(5788), 230–235.CrossRefGoogle Scholar
Hegerl, G.C., Crowley, T.J., Hyde, W.T. and Frame, D.J.. 2006. Climate sensitivity constrained by temperature reconstructions over the past seven centuries. Nature, 440(7087), 1029–1032.Google Scholar
Kondo, O., Miyata, H. and Tôei, K.. 1982. Determination of sulfate in river water by flow injection analysis. Anal. Chim. Acta, 134, 353–358.Google Scholar
Legrand, M.R. and Delmas, R.J.. 1987. Experimental protocol for the chemical analysis of snow, firn and ice cores. In Jones, H.G. and Orville-Thomas, W.J., eds. Seasonal snowcovers: physics, chemistry, hydrology. Dordrecht, etc., D. Reidel Publishing Co., 225–254. (NATO ASI Series C: Mathematical and Physical Sciences 211.)Google Scholar
Legrand, M. and Mayewski, P.. 1997. Glaciochemistry of polar ice cores: a review. Rev. Geophys., 35(3), 219–243.Google Scholar
Littot, G.C. and 9 others. 2002. Comparison of analytical methods used for measuring major ions in the EPICA Dome C (Antarctica) ice core. Ann. Glaciol., 35, 299–305.Google Scholar
Madsen, B.C. and Murphy, R.C.. 1981. Flow injection and photometric determination of sulphate in rainwater with methylthymol blue. Anal. Chem., 53(12), 1924–1926.Google Scholar
Mortensen, A.K., Bigler, M., Grönvold, K., Steffensen, J.P. and Johnsen, S.J.. 2005. Volcanic ash layers from the Last Glacial Termination in the NGRIP ice core. J. Quaternary Sci., 20(3), 209–219.Google Scholar
North Greenland Ice Core Project members. 2004. High-resolution record of Northern Hemisphere climate extending into the last interglacial period. Nature, 431(7005), 147–151.Google Scholar
Osterberg, E.C., Handley, M.J., Sneed, S.B., Mayewski, P.A. and Kreutz, K.J.. 2006. Continuous ice core melter system with discrete sampling for major ion, trace element, and stable isotope analyses. Environ. Sci. Technol., 40(10), 3355–3361.CrossRefGoogle ScholarPubMed
Rasmussen, S.O., Andersen, K.K., Johnsen, S.J., Bigler, M. and McCormack, T.. 2005. Deconvolution-based resolution enhancement of chemical ice core records obtained by continuous flow analysis. J. Geophys. Res., 110, D17304. (10.1029/ 2004JD005717.)Google Scholar
Rasmussen, S.O. and 15 others. 2006. A new Greenland ice core chronology for the last glacial termination. J. Geophys. Res., 111(D06102). (10.1029/2005JD006079.)Google Scholar
Röthlisberger, R. and 6 others. 2000. Technique for continuous high-resolution analysis of trace substances in firn and ice cores. Envir. Sci. Technol., 34(2), 338–342.CrossRefGoogle Scholar
Sigg, A., Fuhrer, K., Anklin, M., Staffelbach, T. and Zurmuehle, D.. 1994. A continuous analysis technique for trace species in ice cores. Environ. Sci. Technol., 28(2), 204–209.CrossRefGoogle ScholarPubMed
Sommer, S., Wagenbach, D., Mulvaney, R. and Fischer, H.. 2000. Glacio-chemical study spanning the past 2 kyr on three ice cores from Dronning Maud Land, Antarctica. 2. Seasonally resolved chemical records. J. Geophys. Res., 105(D24), 29,423–29,433.Google Scholar
Udisti, R. and 6 others. 2000. Holocene electrical and chemical measurements from the EPICA-Dome C ice core. Ann. Glaciol., 30, 20–26.Google Scholar
Wolff, E.W. and 27others. 2006. Southern Ocean sea-ice extent, productivity and iron flux over the past eight glacial cycles. Nature, 440(7083), 491–496.Google Scholar
Yang, Y., Zhang, X.-X. and Korenaga, T.. 1997a. Determination of sulfur dioxide in Southeast Asia by flow injection analysis with on-line reaction column. Anal. Sci., 13 (suppl.), 397–400.Google Scholar
Yang, Y., Zhang, X.-X., Korenaga, T. and Higuchi, K.. 1997b. Determination of passive-sampled sulphur dioxide in ambient air as sulphate ion by flow injection analysis with an in-line reaction column. Talanta, 45(2), 445–450.CrossRefGoogle ScholarPubMed
Zielinski, G.A., Mayewski, P.A., Meeker, L.D., Whitlow, S.I. and Twickler, M.S.. 1996. 110,000-yr record of explosive volcanism from the GISP2 (Greenland) ice core. Quat. Res., 45(2), 109–118.Google Scholar
Figure 0

Fig. 1. Setup of the CFA SO42– detection system. Features include the inline Cation Exchange Column (CEC), the overflow debubbler and the inline BaSO4 reactor column. Flow rates are given for each pump tube of the peristaltic pump. The broken line indicates the fibre optic cables from the flow cell to the lamp and to the detector, respectively.

Figure 1

Fig. 2. (a) Spectrometer signal S (number of counts) of a SO42– standard series (dark grey) with given concentration values. The baseline (light grey) is calculated by linearly interpolating between two sections of blank measurements (black) before and after the standards. (b) Calculated absorption values A of the standard series (dark grey) with respect to the zero line (light grey). Plateau values used to calculate the calibration curve (Fig. 3) are highlighted in black. The insertion shows the measured absorption spectrum A (in arbitrary units) with a maximum around the wavelength λ = 659 nm.

Figure 2

Fig. 3. Calibration curve between measured absorption values and given standard concentrations (points). To calculate the linear fit (A = –0.000 103 C) the two highest values were excluded to account for the deviation from linearity at high concentration levels. The insertion contains an enlargement of the lowest concentration values showing a satisfactory linearity (error bars correspond to the standard deviations of the standard plateaus in Fig. 2).