Hostname: page-component-848d4c4894-4hhp2 Total loading time: 0 Render date: 2024-06-03T09:49:16.765Z Has data issue: false hasContentIssue false

Scintillator - Photodiode Linear Arrays for X-Ray Inspection System

Published online by Cambridge University Press:  10 February 2011

S. M. Ignatov
Affiliation:
RECOM Ltd, Kurchatov Institute,Stchukinskaya Str. 12-1, Moscow, 123182,Russia
V.N. Potapov
Affiliation:
RECOM Ltd, Kurchatov Institute,Stchukinskaya Str. 12-1, Moscow, 123182,Russia
A. V. Fedin
Affiliation:
RECOM Ltd, Kurchatov Institute,Stchukinskaya Str. 12-1, Moscow, 123182,Russia
V. M. Chirkin
Affiliation:
RECOM Ltd, Kurchatov Institute,Stchukinskaya Str. 12-1, Moscow, 123182,Russia
L. I. Urutskoev
Affiliation:
RECOM Ltd, Kurchatov Institute,Stchukinskaya Str. 12-1, Moscow, 123182,Russia
V. V. Gostilo
Affiliation:
Baltic Scientific Instruments, Ganibu Dambis 26, P.O.Box 33, Riga, LV-1005, Latvia
V. V. Kondrashov
Affiliation:
Baltic Scientific Instruments, Ganibu Dambis 26, P.O.Box 33, Riga, LV-1005, Latvia
A. D. Sokolov
Affiliation:
Baltic Scientific Instruments, Ganibu Dambis 26, P.O.Box 33, Riga, LV-1005, Latvia
Get access

Abstract

The evaluations of the following performances of detectors for design of the radiographic system were carried out by mathematical simulation method and experimentally: sensitivity to absorbed energy, noise level, form of apparatus function for spatial resolution of detectors array.

The results of calculations were used for designing of CdWO4-Si arrays with optimal sizes of scintillator and photodiode for resolution ability. At the given stage the quality of radiation detectors was evaluated by spectroscopic methods. At the bias voltage on photodiode 24 V, the sensitivity of detectors has made up (0,95–1,00) × 10−21 C / eV, noise level - (1,50–1,75) ×10−16 C together with electronic noise. Dynamic range at absorbed energy by separate detector is more than 105, what is sufficient to get required image contrast.

Type
Research Article
Copyright
Copyright © Materials Research Society 1998

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

[1]. METORSCAN. Leaflet of METOREX International Oy, 1997.Google Scholar