Chapter 9 - Imaging perfusion and blood–brain barrier permeability using T1-weighted dynamic contrast-enhanced MR imaging

from Section 1 - Physiological MR techniques

Published online by Cambridge University Press:  05 March 2013

Jonathan H. Gillard
Affiliation:
University of Cambridge
Adam D. Waldman
Affiliation:
Imperial College London
Peter B. Barker
Affiliation:
The Johns Hopkins University School of Medicine
Get access

Summary

Introduction

T1-weighted dynamic contrast-enhanced (DCE) magnetic resonance imaging (MRI) is an experimental technique designed to measure tissue hemodynamic parameters.[1–3] For this purpose, contrast medium (the tracer) is rapidly injected intravenously. The passage of the bolus through the tissue causes changes in the longitudinal relaxation rate R1 (1/T1) that are measured with a dynamic T1-weighted imaging technique (Fig. 9.1). The principles of MRI signal theory are used to determine tracer concentration–time curves from the measured signal changes. In a second step, tracer kinetic principles [4,5] are applied to derive the relevant perfusion and/or blood–brain barrier (BBB) permeability parameters.

Perfusion and permeability parameters

Perfusion parameters characterize the state of the microvasculature (Fig. 9.2). Cerebral blood flow (CBF) quantifies tissue perfusion and is conventionally expressed in units of milliliters of blood per minute per 100 g tissue. Cerebral blood volume (CBV) quantifies the volume of the microvasculature and has the units of milliliters of blood per 100 g tissue. Compared with other organs, normal brain tissue is characterized by low blood volumes, with CBV in the range 2–4 ml/100 g.[6]

Type
Chapter
Information
Clinical MR Neuroimaging
Physiological and Functional Techniques
, pp. 113 - 128
Publisher: Cambridge University Press
Print publication year: 2009

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Barbier, EL, Lamalle, L, Décorps, M. Methodology of brain perfusion imaging. J Magn Reson Imaging 2001; 13: 496–520.CrossRefGoogle ScholarPubMed
Tofts, P.Quantitative MRI of the Brain: Measuring Changes Caused by Disease. Chichester, UK: John Wiley, 2003.CrossRefGoogle Scholar
Jackson, A, Buckley, DL, Parker, GJM.Dynamic Contrast-enhanced Magnetic Resonance Imaging in Oncology. Berlin: Springer, 2005.CrossRefGoogle Scholar
Lassen, NA, Perl, WP.Tracer Kinetic Methods in Medical Physiology. New York: Raven Press, 1979.Google Scholar
Jacquez, JA.Compartmental Analysis in Biology and Medicine. Michigan: University of Michigan Press, 1985.Google Scholar
Bassingthwaite, JB.Microcirculatory considerations in NMR flow imaging. Magn Reson Med 1990; 14: 172–178.CrossRefGoogle Scholar
Sorensen, AG, Reimer, P.Cerebral MR Perfusion Imaging. Stuttgart: Georg Thieme, 2000.Google Scholar
Sourbron, S, Michaely, HJ, Reiser, MF, Schoenberg, SO. MRI-measurement of perfusion and glomerular filtration in the human kidney with a separable compartment model. Invest Radiol 2008; 43: 40–48.CrossRefGoogle ScholarPubMed
Buckley, DL, Roberts, C, Parker, GJM, et al. Prostate cancer: evaluation of vascular characteristics with dynamic contrast-enhanced T1-weighted MR imaging – initial experience. Radiology 2006; 233: 709–715.CrossRefGoogle Scholar
Brix, G, Kiessling, F, Lucht, R, et al. Microcirculation and microvasculature in breast tumors: pharmacokinetic analysis of dynamic MR image series. Magn Reson Med 2004; 52: 420–429.CrossRefGoogle ScholarPubMed
Bazelaire, C, Siauve, N, Fournier, L, et al. Comprehensive model for simultaneous MRI determination of perfusion and permeability using a blood-pool agent in rats rhabdomyosarcoma. Eur Radiol 2005; 15: 2497–2505.CrossRefGoogle ScholarPubMed
Larsson, HB, Stubgaard, M, Frederiksen, JL, et al. Quantitation of blood–brain barrier defect by magnetic resonance imaging and gadolinium-DTPA in patients with multiple sclerosis and brain tumors. Magn Reson Med 1990; 16: 117–131.CrossRefGoogle ScholarPubMed
Tofts, PS, Kermode, AG. Measurement of the blood–brain barrier permeability and leakage space using dynamic MR imaging. 1. Fundamental concepts. Magn Reson Med 1991; 40: 229–235.Google Scholar
Covarrubias, DJ, Rosen, BR, Lev, MH. Dynamic magnetic resonance perfusion imaging of brain tumors. Oncologist 2004; 9: 528–537.CrossRefGoogle ScholarPubMed
Kiselev, VG. Transverse relaxation effect of MRI contrast agents: a crucial issue for quantitative measurements of cerebral perfusion. J Magn Reson Imaging 2005; 22: 693–696.CrossRefGoogle ScholarPubMed
Rohrer, M, Bauer, H, Mintorovitch, J, et al. Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Invest Radiol 2005; 40: 715–724.CrossRefGoogle ScholarPubMed
Villringer, A, Rosen, BR, Belliveau, JW, et al. Dynamic imaging with lanthanide chelates in normal brain: contrast due to magnetic susceptibility effects. Magn Reson Med 1988; 6: 164–174.CrossRefGoogle ScholarPubMed
Engvall, C, Ryding, E, Wirestam, R, et al. Human cerebral blood volume (CBV) measured by dynamic susceptibility contrast MRI and 99mTc-RBC SPECT. J Neurosurg Anesthesiol 2008; 20: 41–44.CrossRefGoogle ScholarPubMed
Martel, AL, Allder, SJ, Delay, GS, et al. Perfusion MRI of infarcted and noninfarcted brain tissue in stroke: a comparison of conventional hemodynamic imaging and factor analysis of dynamic studies. Invest Radiol 2001; 36: 378–385.CrossRefGoogle ScholarPubMed
Donahue, KM, Burstein, D, Manning, WJ, Gray, ML.Studies of Gd-DTPA relaxivity and proton exchange rates in tissue. Magn Reson Med 1994; 32: 66–76.CrossRefGoogle ScholarPubMed
Pintaske, J, Martirosian, P, Graf, H, et al. Relaxivity of gadopentetate dimeglumine (magnevist), gadobutrol (gadovist) and gadobenate dimeglumine (multihance) in human blood plasma at 0.2, 1.5 and 3 Tesla. Invest Radiol 2006; 41: 213–221.CrossRefGoogle ScholarPubMed
Dean, BL, Lee, C, Kirsch, JE, et al. Cerebral hemodynamics and cerebral blood volume: MR assessment using gadolinium contrast agents and T1-weighted Turbo-FLASH imaging. AJNR Am J Neuroradiol 1992; 13: 39–48.Google ScholarPubMed
Hackländer, T, Reichenbach, JR, Hofer, M, Mödder, U. Measurement of cerebral blood volume via the relaxing effect of low-dose gadopentetate dimeglumine during bolus transit. AJNR Am J Neuroradiol 1996; 17: 821–830.Google ScholarPubMed
Hackländer, T, Hofer, M, Reichenbach, JR, et al. Cerebral blood volume maps with dynamic contrast-enhanced T1-weighted FLASH imaging: normal values and preliminary clinical results. J Comput Assist Tomogr 1996; 20: 532–539.CrossRefGoogle ScholarPubMed
Hackländer, T, Reichenbach, JR, Mödder, U. Comparison of cerebral blood volume measurements using the T1 and T2* methods in normal human brains and brain tumors. J Comput Assisted Tomogr 1997; 21: 857–866.CrossRefGoogle ScholarPubMed
Moody, AR, Martel, A, Kenton, A, et al. Contrast-reduced imaging of tissue concentration and arterial level (CRITICAL) for assessment of cerebral hemodynamics in acute stroke by magnetic resonance. Invest Radiol 2000; 35: 401–411.CrossRefGoogle ScholarPubMed
Larsson, HB, Hansen, AE, Berg, HK, et al. Dynamic contrast-enhanced quantitative perfusion measurement of the brain using T1-weighted MRI at 3 T. J Magn Reson Imaging 2008; 27: 754–762.CrossRefGoogle Scholar
Bruening, R, Kwong, KK, Vevea, MJ, et al. Echo-planar MR determination of relative cerebral blood volume in human brain tumors: T1 versus T2 weighting. AJNR Am J Neuroradiol 1996; 17: 831–840.Google ScholarPubMed
Sourbron, S, Dujardin, M, Luypaert, R, et al. Quantification of cerebral tumor perfusion and permeability with a deconvolution analysis of T1-weighted bolus tracking data. Magn Reson Mater Phy 2004; 17(Suppl 1): 188.Google Scholar
Pauliah, M, Saxena, V, Haris, M, et al. Improved T1-weighted dynamic contrast-enhanced MRI to probe microvascularity and heterogeneity of human glioma. Magn Reson Imaging 2007; 25: 1292–1299.CrossRefGoogle Scholar
Li, KL, Zhu, X, Waterton, J, Jackson, A.Improved 3D quantitative mapping of blood volume and endothelial permeability in brain tumors. J Magn Reson Imaging 2000; 12: 347–357.3.0.CO;2-7>CrossRefGoogle ScholarPubMed
Roberts, HC, Roberts, TPL, Brasch, RC, Dillon, WP. Quantitative measurement of microvascular permeability in human brain tumors achieved using dynamic contrast-enhanced MR imaging: correlation with histologic grade. AJNR Am J Neuroradiol 2000; 21: 891–899.Google ScholarPubMed
Ewing, JR, Knight, RA, Nagaraja, TN, et al. Patlak plots of Gd-DTPA MRI data yield blood–brain transfer constants concordant with those of 14C-sucrose in areas of blood–brain opening. Magn Reson Med 2003; 50: 283–292.CrossRefGoogle ScholarPubMed
Haroon, HA, Buckley, DL, Patankar, TA, et al. A comparison of Ktrans measurements obtained with conventional and first pass pharmacokinetic models in human gliomas. J Magn Reson Imaging 2004; 19: 527–536.CrossRefGoogle ScholarPubMed
Kassner, A, Roberts, T, Taylor, K, et al. Prediction of hemorrhage in acute ischemic stroke using permeability MR imaging. AJNR Am J Neuroradiol 2005; 26: 2213–2217.Google ScholarPubMed
Singh, A, Haris, M, Rathore, D. et al. Quantification of physiological and hemodynamic indices using T1 dynamic contrast-enhanced MRI in intracranial mass lesions. J Magn Reson Imaging 2007; 26: 871–880.CrossRefGoogle Scholar
Sourbron, S, Ingrisch, M, Siefert, A, et al. Quantification of cerebral blood flow, cerebral blood volume and blood–brain-barrier leakage with DCE-MRI. Magn Reson Med 2009; 62: 205–217.CrossRefGoogle ScholarPubMed
Wardlaw, JM, Farrall, A, Armitage, PA et al. Changes in background blood–brain barrier integrity between lacunar and cortical ischemic stroke subtypes. Stroke 2008; 39: 1327–1332.CrossRefGoogle ScholarPubMed
Parker, GJM, Roberts, C, Macdonald, A, et al. Experimentally-derived functional form for a population-averaged high-temporal-resolution arterial input function for dynamic contrast-enhanced MRI. Magn Reson Med 2006; 56: 993–1000.CrossRefGoogle ScholarPubMed
Roberts, C, Issa, B, Stone, A, et al. Comparative study into the robustness of compartmental modeling and model-free analysis in DCE-MRI studies. J Magn Reson Imaging 2006; 23: 554–563.CrossRefGoogle ScholarPubMed
Sourbron, S, Luypaert, R, Morhard, D, et al. Deconvolution of bolus-tracking data: a comparison of discretization methods. Phys Med Biol 2007; 52: 6761–6778.CrossRefGoogle ScholarPubMed
Haris, M, Gupta, RK, Singh, A, et al. Differentiation of infective from neoplastic brain lesions by dynamic contrast-enhanced MRI. Neuroradiology 2008; 50: 531–540.CrossRefGoogle ScholarPubMed
Sourbron, S, Dujardin, M, Makkat, S, Luypaert, R. Pixel-by-pixel deconvolution of bolus-tracking data: optimization and implementation. Phys Med Biol 2007; 52: 429–447.CrossRefGoogle ScholarPubMed
Østergaard, L, Weisskoff, RM, Chesler, DA, et al. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: mathematical approach and statistical analysis. Magn Reson Med 1996; 36: 715–725.CrossRefGoogle ScholarPubMed
Tofts, PS, Brix, G, Buckley, DL, et al. Estimating kinetic parameters from dynamic contrast-enhanced T1-weighted MRI of a diffusible tracer: standardized quantities and symbols. J Magn Reson Imaging 1999; 102: 223–232.3.0.CO;2-S>CrossRefGoogle Scholar
Rijpkema, M, Kaanders, JHAM, Joosten, FBM, et al. Method for quantitative mapping of dynamic MRI contrast agent uptake in human tumors. J Magn Reson Imaging 2001; 14: 457–463.CrossRefGoogle ScholarPubMed
Harris, NG, Gauden, V, Fraser, PA, et al. MRI measurement of blood–brain barrier permeability following spontaneous reperfusion in the starch microsphere model of ischemia. Magn Reson Imaging 2000; 20: 221–230.CrossRefGoogle Scholar
Jackson, A, Jayson, GC, Li, KL, et al. Reproducibility of quantitative dynamic contrast-enhanced MRI in newly presenting glioma. Br J Radiol 2003; 76: 153–162.CrossRefGoogle ScholarPubMed
Harrer, J, Parker, G, Haroon, H, et al. Comparative study of methods for determining vascular permeability and blood volume in human gliomas. J Magn Reson Imaging 2004; 20: 748–757.CrossRefGoogle ScholarPubMed
Armitage, P, Schwindack, C, Bastin, M, Whittle, I. Quantitative assessment of intracranial tumor response to dexamethasone using diffusion, perfusion and permeability magnetic resonance imaging. Magn Reson Imaging 2007; 25: 303–310.CrossRefGoogle ScholarPubMed
Ewing, JR, Brown, SL, Nagaraja, TN et al. MRI measurement of change in vascular parameters in the 9L rat cerebral tumor after dexamethasone administration. J Magn Reson Imaging 2008; 27: 1430–1438.CrossRefGoogle ScholarPubMed
Patlak, CS, Blasberg, RG, Fenstermacher, JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. J Cereb Blood Flow Metab 1983; 3: 1–7.CrossRefGoogle ScholarPubMed
Taheri, S, Sood, R. Kalman filtering for reliable estimation of BBB permeability. Magn Reson Imaging 2006; 24: 1039–1049.CrossRefGoogle ScholarPubMed
Heiland, S, Benner, T, Debus, J, et al. Simultaneous assessment of cerebral hemodynamics and contrast agent uptake in lesions with disrupted blood–brain-barrier. Magn Reson Imaging 1999; 17: 21–27.CrossRefGoogle ScholarPubMed
Uematsu, H, Maeda, M. Double-echo perfusion-weighted MR imaging: basic concepts and application in brain tumors for the assessment of tumor blood volume and vascular permeability. Eur Radiol 2006; 16: 180–186.CrossRefGoogle ScholarPubMed
Sourbron, S, Heilmann, M, Biffar, A, et al. Bolus-tracking MRI with a simultaneous T1- and T2*-measurement. Magn Reson Med 2009; Epub ahead of print; DOI 10.1002/mrm.22042.
Landis, CS, Li, X, Telang, FW. Equilibrium transcytolemmal water-exchange kinetics in skeletal muscle in vivo. Magn Reson Med 1999; 42: 467–478.3.0.CO;2-0>CrossRefGoogle ScholarPubMed
Larsson, HBW, Rosenbaum, S, Fritz-Hansen, T. Quantification of the effect of water exchange in dynamic contrast MRI perfusion measurements in the brain and heart. Magn Reson Med 2001; 46: 272–281.CrossRefGoogle ScholarPubMed
Buckley, DL. Transcytolemmal water exchange and its affect on the determination of contrast agent concentration in vivo. Magn Reson Med 2002; 47: 420–421.CrossRefGoogle ScholarPubMed
Buckley, DL, Kershaw, LE, Stanisz, G. Cellular–interstitial water exchange and its effect on the determination of contrast agent concentration in vivo: dynamic contrast-enhanced MRI of human internal obturator muscle. Magn Reson Med 2008; 60: 1011–1019.CrossRefGoogle ScholarPubMed
Donahue, KM, Weisskoff, RM, Chesler, DA, et al. Improving MR quantification of regional blood volume with intravascular T1 contrast agents: Accuracy, precision and water exchange. Magn Reson Med 1996; 36: 858–867.CrossRefGoogle ScholarPubMed

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×