Semi-automated speckle-tracking for quantitative right ventricular assessment in children and adolescents

  • Aron K. McCloud (a1) (a2), Joanna Lowisz (a1), David A. Roberson (a1), Cheryl A. Lefaiver (a3) and Jamie S. Penk (a1)...

Abstract

Background:

Assessment of right ventricular size and function is an important part of the clinical cardiac evaluation; however, these quantitative measures are challenging by echocardiography. Automated software could be useful in place of manual measurements and qualitative assessment. This study evaluates a semi-automated software by comparing automated to manual measures in normal children.

Methods:

Neonates to adolescents with normal echocardiograms were prospectively enrolled. Measurements were performed using manual techniques and semi-automated software (EchoInsight®, Epsilon Imaging, Ann Arbor, Michigan, United States of America). Right ventricular measurements included end-diastolic and end-systolic area, fractional area change, chamber dimensions, and tricuspid annular plane systolic excursion. Agreement between manual and semi-automated measures was compared.

Results:

Echocardiograms for 233 patients were included in the analysis. Intra- and inter-observer reliabilities for semi-automated measures were good with intraclass correlation coefficients all over 0.9 and 0.85, respectively. There was very strong correlation between manual and semi-automated methods for areas and dimensions (r = 0.93–0.99) and low bias (1.4–10.8%). For functional measures, tricuspid annular plane systolic excursion measures correlated well (r = 0.84), but fractional area change did not (r = 0.50). Both demonstrated significant bias (33.5–43.0%). The semi-automated method consistently underestimated fractional area change with a mean of 26.6% versus a manual mean of 36.1%.

Conclusions:

The semi-automated software is capable of generating quantitative right ventricular measures in children with good reliability. The software demonstrates very good correlation and low bias when compared to manual methods for right ventricular areas and dimensions. There is a significant difference between manual and semi-automated techniques for the functional measures.

Copyright

Corresponding author

Author for correspondence: A. K. McCloud, DO, Division of Pediatric Cardiology, University of Virginia, 629 Cedar Creek Grade, Suite B, Winchester, VA 22601, USA. Tel: (540) 678-3950; Fax: (540) 678-3954; E-mail: akm6v@virginia.edu

Footnotes

Hide All
*

Presented at the 21st Annual Update on Pediatric and Congenital Cardiovascular Disease, Scottsdale, AZ, USA, 23 February 2018.

Footnotes

References

Hide All
1. Lai, WW, Geva, T, Shirali, GS, et al. Guidelines and standards for performance of a pediatric echocardiogram: a report from the Task Force of the Pediatric Council of the American Society of Echocardiography. J Am Soc Echocardiogr 2006; 19: 14131430. doi: 10.1016/j.echo.2006.09.001.
2. Lopez, L, Colan, SD, Frommelt, PC, et al. Recommendations for quantification methods during the performance of a pediatric echocardiogram: a report from the Pediatric Measurements Writing Group of the American Society of Echocardiography Pediatric and Congenital Heart Disease Council. J Am Coll Cardiol 2010; 23: 465495. doi: 10.1016/j.echo.2010.03.019.
3. Saguner, AM, Vecchiati, A, Baldinger, SH, et al. Different prognostic value of functional right ventricular parameters in arrhythmogenic right ventricular cardiomyopathy/dysplasia. Circ Cardiovasc Imaging 2014; 7: 230239. doi: 10.1161/CIRCIMAGING.113.000210.
4. Leong, DP, Höke, U, Delgado, V, et al. Right ventricular function and survival following cardiac resynchronisation therapy. Heart 2013; 99: 722728. doi: 10.1136/heartjnl-2012-303076.
5. Mahle, WT, Coon, PD, Wernovsky, G, et al. Quantitative echocardiographic assessment of the performance of the functionally single right ventricle after the Fontan operation. Cardiol Young 2001; 11: 399406.
6. Davlouros, PA, Niwa, K, Webb, G, et al. The right ventricle in congenital heart disease. Heart 2006; 92 (Suppl 1): i27i38. doi: 10.1136/hrt.2005.077438.
7. Lang, RM, Bierig, M, Devereux, RB, et al. Recommendations for chamber quantification: a report from the American Society of Echocardiography’s Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Associateion of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiogr 2005; 18: 14401463. doi: 10.1016/j.echo.2005.10.005.
8. Mor-Avi, V, Lang, RM, Badano, LP, et al. Current and evolving echocardiographic techniques for the quantitative evaluation of cardiac mechanics: ASE/EAE consensus statement on methodology and indications endorsed by the Japanese Society of Echocardiography. J Am Soc Echocardiogr 2011; 24: 277313. doi: 10.1016/j.echo.2011.01.015.
9. Van der Zwaan, HB, Geleijnse, ML, McGhie, JS, et al. Right ventricular quantification in clinical practice: two-dimensional vs. three-dimensional echocardiography compared with cardiac magnetic resonance imaging. Eur J Echocardiogr 2011; 12: 656664. doi: 10.1093/ejechocard/jer107.
10. Bleeker, GB, Steendijk, P, Holman, ER, et al. Assessing right ventricular function: the role of echocardiography and complementary technologies. Heart 2006; 92 (Suppl 1): i19i26. doi: 10.1136/hrt.2005.082503.
11. Wang, J, Prakasa, K, Bomma, C, et al. Comparison of novel echocardiographic parameters of right ventricular function with ejection fraction by cardiac magnetic resonance. J Am Soc Echocardiogr 2007; 20: 10581064. doi: 10.1016/j.echo.2007.01.038.
12. Tamborini, G, Pepi, M, Galli, CA, et al. Feasibility and accuracy of a routine echocardiographic assessment of right ventricular function. Int J Cardiol 2007; 115: 8689. doi: 10.1016/j.ijcard.2006.01.017.
13. Lai, WW, Gauvreau, K, Rivera, ES, et al. Accuracy of guideline recommendations for two-dimensional quantification of the right ventricle by echocardiography. Int J Cardiovasc Imaging 2008; 24: 691698. doi: 10.1007/s10554-008-9314-4.
14. Kopecna, D, Briongos, S, Castillo, H, et al. Interobserver reliability of echocardiography for prognostication of normotensive patients with pulmonary embolism. Cardiovasc Ultrasound 2014; 12: 29. doi: 10.1186/1476-7120-12-29.
15. Michalis, LK, Thomas, MR, Jewitt, DE, et al. Echocardiographic assessment of systolic and diastolic left ventricular function using an automatic boundary detection system. Correlation with established invasive and non invasive parameters. Int J Card Imaging 1995; 11: 7180. doi: 10.1007/BF01844704
16. Yagi, T, Yoshida, K, Hozumi, T, et al. Automatic assessment of left ventricular cavity area by the automated contour tracking method. J Cardiol 1996; 28: 345348.
17. Sugioka, K, Hozumi, T, Yagi, T, et al. Automated quantification of left ventricular function by the automated contour tracking method. Echocardiography 2003; 20: 313318. doi: 10.1046/j.1540-8175.2003.03036.
18. Rahmouni, HW, Ky, B, Plappert, T, et al. Clinical utility of automated assessment of left ventricular ejection fraction using artificial intelligence-assisted border detection. Am Heart J 2008; 155: 562570. doi: 10.1016/j.ahj.2007.11.002.
19. Oe, M, Gorcsan, J 3rd, Mandarino, WA, et al. Automated echocardiographic measures of right ventricular area as an index of volume and end-systolic pressure-area relations to assess right ventricular function. Circulation 1995; 92: 10261033.
20. Forni, G, Pozzoli, M, Cannizzaro, G, et al. Assessment of right ventricular function in patients with congestive heart failure by echocardiographic automated boundary detection. Am J Cardiol 1996; 78: 13171321. doi: 10.1016/S0002-9149(96)00623-6.
21. Greiner, S, André, F, Heimisch, M, et al. Non-invasive quantification of right ventricular systolic function by echocardiography: a new semi-automated approach. Clin Res Cardiol 2013; 102: 229235. doi: 10.1007/s00392-012-0528-z.
22. Kimball, TR, Witt, SA, Khoury, PR, et al. Automated echocardiographic analysis of systemic ventricular performance in hypoplastic left heart syndrome. J Am Soc Echocardiogr 1996; 9: 629636.
23. Helbing, WA, Bosch, HG, Maliepaard, C, et al. On-line automated border detection for echocardiographic quantification of right ventricular size and function in children. Pediatr Cardiol 1997; 18: 261699. doi: 10.1007/s002469900171.
24. Gao, Z, Bortman, J, Mahmood, F, et al. Vendor-neutral right ventricular strain measurement. J Cardiothorac Vasc Anesth 2018; 32: 17591767. doi: 10.1053/j.jvca.2018.02.014.
25. Medvedofsky, D, Addetia, K, Hamilton, J, et al. Semi-automated echocardiographic quantification of right ventricular size and function. Int J Cardiovasc Imaging 2015; 31: 11491157. doi: 10.1007/s10554-015-0672-4.
26. Bland, M, Altman, DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol 2003; 22: 8593. doi: 10.1002/uog.122.
27. Knight, DS, Schwaiger, JP, Krupickova, S, et al. Accuracy and test-retest reproducibility of two-dimensional knowledge-based volumetric reconstruction of the right ventricle in pulmonary hypertension. J Am Soc Echocardiogr 2015; 28: 989998. doi: 10.1016/j.echo.2015.02.020.

Keywords

Type Description Title
VIDEO
Supplementary materials

McCloud et al. supplementary material
McCloud et al. supplementary material 1

 Video (32.1 MB)
32.1 MB

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed