Cleaning of Blood-Contaminated Reprocessed Angiographic Catheters and Spinal Needles

  • Thereza Christina Vessoni Penna (a1) and Carlos Augusto M. Ferraz (a2)
Abstract
Objectives:

To evaluate the efficacy of a multistep cleaning method using a cleaner and a chemical disinfectant on blood-contaminated angiographic catheters and spinal needles intended to be sterilized by hydrogen peroxide gas plasma.

Method:

A mixture of radiopaque iodine contrast, bovine blood (plus ethylenediaminetetraacetic acid), and a suspension of Bacillus subtilis spores was used to simulate catheterization and needle use. The mixture was a 1:1 proportion of contrast and blood, inoculated so that there was a final concentration of B subtilis spores of 1.0×106 colony-forming units (CFU)/mL. The inoculated devices were cleaned using a hydrogen peroxide solution at a concentration of 1.5±0.5 percent by weight, followed by distilled water with enzymatic detergent. After drying, the devices were sterilized with hydrogen peroxide gas plasma.

Results:

The initial B subtilis spore concentration inoculated into catheters and needles varied from 2.12×104 to 2.74×107 CFU/mL. The residual load of B subtilis spores after cleaning varied from zero (no count) to a maximum of 200 CFU/device. The multistep cleaning procedure was responsible for an average 5-log10 reduction of B subtilis spores in the catheter and needle lumens.

Conclusions:

The hydrogen peroxide and enzymatic detergent aqueous solutions were shown to be efficacious when used as part of a multistep cleaning method. The low level of microbial contamination prior to sterilization with hydrogen peroxide gas plasma assured that the intended sterility assurance level was reached.

Copyright
Corresponding author
Rua Antonio de Macedo Soares, 452, São Paulo, SP 04607-000, Brasil
References
Hide All
1.Association for the Advancement of Medical Instrumentation (US). Design, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers. Report no. 12. Arlington, VA: AAMI; 1995.
2.Rutala, WA. Disinfection and sterilization of patient-care items. Infect Control Hosp Epidemiol 1996;17:377384.
3.Russel, D, Furr, JR, Maillard, JY. Microbial susceptibility and resistance to biocides. ASM News 1997;63:481487.
4.Alfa, MJ, DeGagne, P, Olson, N, Puchalski, T. Comparison of ion plasma, vaporized hydrogen peroxide, and 100% ethylene oxide sterilizers to the 12/88 ethylene oxide gas sterilizer. Infect Control Hosp Epidemiol 1996;17:92100.
5.Bentolita, P, Jacob, R, Roberge, F. Effects of re-use on the physical characteristics of angiographic catheters. J Med Eng Technol 1990;4:254259.
6.Jacobson, JA, Schwartz, CE, Marshall, HW, Conti, M, Burke, JP. Fever, chills, and hypotension following cardiac catheterization with single- and multiple-use disposable catheters. Cathet Cardiovasc Diagn 1983;9:3946.
7.American Public Health Association. Standard Methods for the Examination of Water and Wastewater. 19th ed. Washington, DC: APHA; 1995:938.
8.Penna, TCV, Ferraz, CAM, Cassola, MA. Presterilization microbial load on reused medical devices and STERRAD 100 effectiveness on reduction of inoculated Bacillus subtilis spores. Infect Control Hosp Epidemiol 1999;20:465472.
9.Cassola, MA, Ishisaki, E, Ferraz, CA, Penna, TCV. Hydrogen peroxide gas plasma sterilization of medical devices in Brazilian hospitals. Laes & Haes 1997;19:88100.
10.United States Pharmacopeia. 23rd ed. Rockville, MD: United States Pharmacopeial Convention; 1995:1847-1849,19761981.
11.Myers, HC, McAlister, D, Antonoplos, P. Natural bioburden levels detected on rigid lumened medical devices before and after cleaning. Am J Infect Control 1997;25:471476.
12.Chu, NS, Myers, HC, Ghazanfari, N, Antonoplos, P. Levels of naturally occurring microorganisms on surgical instruments after use and after washing. Am J Infect Control 1999;27:315319.
13.Rutala, WA, Gerden, MF, Jones, JF, Weber, DJ. Levels of microbial contamination on surgical instruments. Am J Infect Control 1998;26:143145.
14.Alfa, MJ, Degagne, P, Olson, N. Worst-case soiling levels for patient-used flexible endoscopes before and after cleaning. Am J Infect Control 1999;27:392401.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Infection Control & Hospital Epidemiology
  • ISSN: 0899-823X
  • EISSN: 1559-6834
  • URL: /core/journals/infection-control-and-hospital-epidemiology
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 3 *
Loading metrics...

Abstract views

Total abstract views: 73 *
Loading metrics...

* Views captured on Cambridge Core between September 2016 - 15th August 2018. This data will be updated every 24 hours.