Skip to main content
×
×
Home

Autofocus by Bayes Spectral Entropy Applied to Optical Microscopy

  • Steffen Podlech (a1)
Abstract

This study introduces a passive autofocus method based on image analysis calculating the Bayes spectral entropy (BSE). The method is applied to optical microscopy and together with the specific construction of the opto-mechanical unit, it allows the analysis of large samples with complicated surfaces without subsampling. This paper will provide a short overview of the relevant theory of calculating the normalized discrete cosine transform when analyzing obtained images, in order to find the BSE measure. Furthermore, it will be shown that the BSE measure is a strong indicator, helping to determine the focal position of the optical microscope. To demonstrate the strength and robustness of the microscope system, tests have been performed using a 1951 USAF test pattern resolution chart determining the in focus position of the microscope. Finally, this method and the optical microscope system is applied to analyze an optical grating (100 lines/mm) demonstrating the detection of the focal position. The paper concludes with an outlook of potential applications of the presented system within quality control and surface analysis.

Copyright
Corresponding author
*Corresponding author.stp@rybners.dk
References
Hide All
Geusebroek, J.-M., Cornelissen, F., Smeulders, A.W.M. & Geerts, H. (2000). Robust autofocussing in microscopy. Cytometry 39, 19.
Gonzalez, R.C. & Woods, R.E. (2008). Digital Image Processing. Upper Saddle River, NJ, USA: Pearson Prentice Hall.
Hotchkiss, R.N., Washer, F.E. & Rosberry, F.W. (1951). Spurious resolution of photographic lenses. J Acoust Soc Am 41, 600603.
Jeffreys, S.H. & Swirles, B. (1966). Methods of Mathematical Physics. Cambridge, UK: Cambridge University Press.
Kristan, M., Perš, J., Perše, M. & Kovačič, S. (2006). Bayes-spectral entropy-based measure of camera focus using a discrete cosine transform. Pattern Recognit Lett 27, 14311439.
Neumann, B., Dämon, A., Hogenkamp, D., Beckmann, E. & Kollmann, J. (1989). A laser-autofocus for automatic microscopy and metrology. Sensors Actuators 17, 267272.
Santoz, A., Ortiz de Solórzano, C., Vaquero, J.J., Peña, J. & Malpica, N. (1997). Evaluation of autofocus functions in molecular cytogenetic analysis. J Microsc 188, 264272.
Stewart, J.E. (1965). Spurious spectral resolution and the spectral transfer function of monochromators. Appl Opt 4(5), 609612.
Vollath, D. (1988). Automatic focussing by correlative methods. J Microsc 147, 279288.
Xu, X., Wang, Y., Tang, J., Zhang, X. & Liu, X. (2011). Robust automatic focus algorithm for low contrast images using a new contrast measure. Sensors 11, 82818294.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Microscopy and Microanalysis
  • ISSN: 1431-9276
  • EISSN: 1435-8115
  • URL: /core/journals/microscopy-and-microanalysis
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Keywords

Metrics

Full text views

Total number of HTML views: 8
Total number of PDF views: 49 *
Loading metrics...

Abstract views

Total abstract views: 294 *
Loading metrics...

* Views captured on Cambridge Core between September 2016 - 15th August 2018. This data will be updated every 24 hours.