Some Advantages of EDS Analysis in a 400 KV Electron Microscope
Published online by Cambridge University Press: 25 February 2011
Abstract
The dependence of the characteristic and bremsstrahlung X-ray counts, the peak to background (P/B) ratio and the spatial resolution on the incident beam energy between 100 keV and 400 keV were measured using a high voltage electron microscope (HVEM). The bremsstrahlung count decreases much faster than that of the characteristic count with the increase of the incident beam energy. The decrease rate depends on Z number. It is ascertained that the P/B ratio and the spatial resolution at 400 keV were 2 or 3 and 2.5 times better than those at 100 keV, respectively.