Case 48 - Pseudothrombus on dark blood images

from Section 6 - Cardiovascular MRI artifacts

Published online by Cambridge University Press:  05 June 2015

Stefan L. Zimmerman
Affiliation:
Johns Hopkins Medical Centre
Stefan L. Zimmerman
Affiliation:
Johns Hopkins Medical Centre
Elliot K. Fishman
Affiliation:
Johns Hopkins Medical Centre
Get access

Summary

Imaging description

High-signal mimicking thrombus can occur in cardiac chambers on inversion recovery-based dark blood images due to slow flow. Double inversion recovery dark blood images use an initial non-slice-selective 180 degree inversion pulse followed immediately by a second slice-selective inversion pulse in the plane of interest. The net effect is to invert all protons outside of the imaging plane while leaving protons in the plane unaffected. Image acquisition begins when inverted blood protons cross the null point during T1 recovery, typically corresponding to mid-diastole. The sequence relies on flowing blood to replace non-inverted blood protons in the image plane with nulled blood from outside the plane. If there is slow moving blood in a cardiac chamber, there will be incomplete blood suppression. In patients with localized hypokinetic regions, artifactual high signal in these regions of stasis is common and can mimic thrombus (Figure 48.1). This is also common in the subendocardial region in patients with reduced ejection fraction due to slow flow in trabeculations (Figure 48.2).

Importance

Misdiagnosis of thrombus in cardiac chambers could lead to risks from anticoagulation and additional unnecessary followup imaging.

Typical clinical scenario

Artifactual signal from slow flow on dark blood images is common in the left ventricular apex, both in the setting of a global cardiomyopathy or prior apical myocardial infarction. However, artifactual high signal can be seen anywhere there is sluggish flow, including the right ventricle and atria.

Differential diagnosis

Pseudothrombus on dark blood images should be distinguished from a true thrombus. Both of these entities can occur in the setting of blood stasis. However, unlike dark blood artifacts, true thrombus should be present on all pulse sequences both pre- and post-contrast (Figure 48.1).

Type
Chapter
Information
Pearls and Pitfalls in Cardiovascular Imaging
Pseudolesions, Artifacts, and Other Difficult Diagnoses
, pp. 150 - 153
Publisher: Cambridge University Press
Print publication year: 2015

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Saremi, F., Grizzard, J. D., Kim, R. J.. Optimizing cardiac MR imaging: practical remedies for artifacts. Radiographics 2008; 28: 1161–87.CrossRefGoogle ScholarPubMed

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×