Case 50 - Aliasing artifact in phase-contrast cardiac MR

from Section 6 - Cardiovascular MRI artifacts

Published online by Cambridge University Press:  05 June 2015

Stefan L. Zimmerman
Affiliation:
Johns Hopkins Medical Centre
Stefan L. Zimmerman
Affiliation:
Johns Hopkins Medical Centre
Elliot K. Fishman
Affiliation:
Johns Hopkins Medical Centre
Get access

Summary

Imaging description

Aliasing artifact occurs in phase-contrast MRI when the peak velocity of flowing blood being imaged exceeds the encoding velocity, or Venc. In phase-contrast MRI, specialized flowsensitive gradients are used to encode flow velocity and flow direction information into grayscale pixel values. The Venc is a parameter, expressed in cm/sec, that specifies the maximal velocity that can be measured by a given phase-contrast acquisition. The Venc is adjustable and is set at the MRI scanner before image aquisition. When the vessel being imaged contains flowing blood that is moving faster than the Venc, aliasing will occur. Pixels will become progressively brighter as velocities approach the Venc; however, pixels that represent velocities exceeding the Venc wrap around and are mapped to dark pixels from the opposite end of the grayscale spectrum (Figure 50.1). Depending on the direction of flow, the inverse situation can also occur, where white aliased pixels may be surrounded by dark non-aliased pixels with velocities just below the Venc (Figure 50.1).

Importance

Aliasing will result in inaccurate measurement of peak velocities within a vessel. Peak velocities are used to estimate the pressure gradient across a stenosis which can dictate treatment decisions. Automated flow measurement software can be used to correct for aliasing if peak velocity is less than three times the Venc. However, repeated acquisitions at an increased Venc setting are preferred for maximum measurement reliability.

Typical clinical scenario

Aliasing artifacts are seen whenever the flow velocity is greater than expected when setting the Venc. This is common in the setting of stenotic vessels or valvular stenosis. Elevated velocities may also be seen within the left ventricular outflow tract in the setting of obstructive hypertrophic cardiomyopathy. Typical maximal velocities for ascending aorta, pulmonary artery, and systemic veins are less than 150 cm/sec, less than 100 cm/sec, and less than 50 cm/sec, respectively. In the setting of stenosis, maximal velocities may increase to 500–1000 cm/sec, and repeated acquisitions with incremental increases of Venc may be necessary to find the optimal Venc to prevent aliasing.

Type
Chapter
Information
Pearls and Pitfalls in Cardiovascular Imaging
Pseudolesions, Artifacts, and Other Difficult Diagnoses
, pp. 159 - 161
Publisher: Cambridge University Press
Print publication year: 2015

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Lotz, J., Meier, C., Leppert, A., Galanski, M.. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics 2002; 22: 651–71.CrossRefGoogle ScholarPubMed
2. Srichai, M. B., Lim, R. P., Wong, S., Lee, V. S.. Cardiovascular applications of phase-contrast MRI. AJR Am J Roentgenol 2009; 192: 662–75.CrossRefGoogle ScholarPubMed

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×