Light Element Analysis with TXRF

Published online by Cambridge University Press:  06 March 2019

Christina Streli
Affiliation:
Atominstitut der Österreichischen Universitäten Schüttelstraβe 115, A-1020 Wien, Austria
Peter Wobrauschek
Affiliation:
Atominstitut der Österreichischen Universitäten Schüttelstraβe 115, A-1020 Wien, Austria
Hannes Aiginger
Affiliation:
Atominstitut der Österreichischen Universitäten Schüttelstraβe 115, A-1020 Wien, Austria
Get access

Abstract

Total Reflection X-Ray Fluorescence Analysis (TXRF) has become a powerful analytical tool for trace element analysis. Because of its advantages in excitation and background reduction TXRF has been applied for the analysis of light elements (C,O,F,Na,...). A special Ge(HP) detector offering an ultra thin window in combination with a spectrometer specially designed for the requirements of light element analysis was used. Also a new windowless X-ray tube for efficient excitation of the light elements was tested. The system was checked with standard aqueous solutions; detection limits in the ng range (7 ng for O) are obtained.