Large-diameter graft-stent (Advanta V12) implantation in various locations: early results

  • Dietmar Schranz (a1), Christian Jux (a1), Melanie Vogel (a1), Jürgen Bauer (a1), Hakan Akintürk (a1) and Klaus Valeske (a1)...
Abstract
AbstractObjectives

Transcatheter stent placement carries the risk of cardiovascular aneurysm or rupture. Covered stent implantation reduces these risks. The recently marketed Advanta V12 large-diameter-covered stent is pre-mounted and requires 9 (8)-11 Fr delivery systems. The aim was to report on the early results of the treatment of various cardiovascular obstructions by the implantation of a new polytetrafluoroethylene-covered stent (V12).

Methods

Graft stents on balloons with a diameter (12, 14, 16 millimetres) sufficient to anchor the stent in various obstructions (congenital aortic coarctation, n = 5; obstruction after ascending aorta repair, n = 2; pulmonary arteries, n = 5; inferior caval vein, n = 1; atretic superior caval vein, n = 1; pulmonary vein obstruction, n = 1; and right ventricular outflow tract, n = 1) were implanted using the smallest available delivery system. Secondary dilation with larger-diameter balloons was performed when the residual pressure was gradient, the stent-vessel wall relationship or stent re-coiling due to different reasons needed a re-intervention by pure ballooning or second stent placement.

Results

All 16 patients aged 5–46 years underwent V12 implantation. The variability of the treated lesions and the need for additional interventions were responsible for large ranges in fluoroscopy time between 7.3 to 48.2 minutes (median 17.3). Considering the additional procedures, the V12 stent achieved the desired result in all cases. There were no major complications. At short-term median follow-up of 2 months, all patients are alive and well with no evidence of stent failing.

Conclusion

These initial results show that the covered Advanta V12 large-diameter stent is safe and effective in the immediate treatment of various cardiovascular obstructions. Long-term follow-up is required.

Copyright
Corresponding author
Correspondence to: Professor Dr D. Schranz, Pediatric Heart Center, Justus-Liebig University, Feulgenstrasse 12, 35385 Giessen, Germany. Tel: 049 641 9943461 Fax: 049 641 9943469; E-mail: Dietmar.Schranz@paediat.med.uni-giessen.de
References
Hide All
1. O’Laughlin MP, Perry SB, lock JE, Mullins CE. Use of endovascular stents in congenital heart disease. Circulation 1991; 83: 19231939.
2. Fogelman R, Nykanen D, Smallhorn JF, et al. Endovascular stents in the pulmonary circulation: clinical impact on management and medium-term follow-up. Circulation 1995; 92: 881885.
3. Ebeid MR, Prieto LR, Latson LA. Use of balloon-expandable stents for coarctation of the aorta: initial results and intermediate-term follow-up. J Am Coll Cardiol 1997; 30: 18471852.
4. Ovaert C, Calderone CA, McCrindle BW, et al. Endovascular stent implantation for the management of postoperative right ventricular outflow tract obstruction: clinical efficacy. J Thorac Cardiovasc Surg 1999; 118: 886893.
5. Ing F. Stents: what’s available to the pediatric interventional cardiologist’s. Catheter Cardiovasc Interv 2002; 57: 374386.
6. Kreutzer J, Rome JJ. Open-cell design stents in congenital heart disease: a comparison of IntraStent vs Palmaz stents. Catheter Cardiovasc Interv 2002; 56: 400409.
7. Hussein H, Walsh K. First in man use of the Advanta trademark V12 e PTFE-coated stent in aortic coarctation. Catheter Cardiovasc Interv 2009; 74: 101102.
8. Bruckheimer E, Birk E, Santiago R, Dagan T, Esteves C, Pedra CAC. Coarctation of the aorta treated with the Advanta V12 large diameter stent: acute results. Catheter Cardiovasc Interv 2010; 75: 402406.
9. Bokenkamp R, Blom N, deWolf D, et al. Intraoperative stenting of pulmonary arteries. Eur J Cardiothorac Surg 2005; 27: 544547.
10. Tomita H, Yazaki S, Kimura K, et al. Potential goals for the dimensions of the pulmonary arteries and aorta with stenting after the Fontan operation. Catheter Cardiovasc Interv 2002; 56: 246253.
11. Bacha EA, Marshall AC, McElhinney DB, del Nido PJ. Expanding the hybrid concept in congenital heart surgery. Semin Thorac Cardiovasc Surg pediatr Card Surg Annu 2007: 146150.
12. Ing FF, Grifka RG, Nihill MR, Mullins CE. Repeat dilatation of intravascular stents in congenital heart defects. Circulation 1995; 92: 893897.
13. Carano N, Agnetti A, Tchana B, Squarcia A, Saquarcia U. Descending thoracic aorta to left pulmonary fistula after stent implantation for acquired left pulmonary artery stenosis. J Interv Cardiol 2002; 15: 11413.
14. Ewert P, Schubert S, Peters B, Abdul-Khaliq H, Nagdyman N, Lange PE. The CP stent – short, long, covered – for the treatment of aortic coarctation, stenosis of pulmonary arteries and caval veins, and Fontan anastomosis in children and adults: an evaluation of 60 stents in 53 patients. Heart 2005; 91: 948953.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Cardiology in the Young
  • ISSN: 1047-9511
  • EISSN: 1467-1107
  • URL: /core/journals/cardiology-in-the-young
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Keywords:

Metrics

Full text views

Total number of HTML views: 1
Total number of PDF views: 7 *
Loading metrics...

Abstract views

Total abstract views: 115 *
Loading metrics...

* Views captured on Cambridge Core between September 2016 - 20th October 2017. This data will be updated every 24 hours.