Penumbra width determination of single beam and 201 beams of Gamma Knife machine model 4C using Monte Carlo simulation

  • Atefeh Mahmoudi (a1), Alireza Shirazi (a1), Ghazale Geraily (a1), Tahereh Hadisi nia (a1), Masoume Bakhshi (a1) and Maryam Maleki (a2)...
Abstract
Background

One of the stereotactic radiosurgery techniques is Gamma Knife radiosurgery, in which intracranial lesions that are inaccessible or inappropriate for surgery are treated using 201 cobalt-60 sources in one treatment session. In this conformal technique, the penumbra width, which results in out-of-field dose in tumour-adjacent normal tissues should be determined accurately. The aim of this study is to calculate the penumbra widths of single and 201 beams for different collimator sizes of Gamma Knife machine model 4C using EGSnrc/BEAMnrc Monte Carlo simulation code and comparison the results with EBT3 film dosimetry data.

Methods and materials

In this study, simulation of Gamma Knife machine model 4C was performed based on the Monte Carlo codes of EGSnrc/BEAMnrc. To investigate the physical penumbra width (80−20%), the single beam and 201 beams profiles were obtained using EGSnrc/DOSXYZnrc code and EBT3 films located at isocentre point in a spherical Plexiglas head phantom.

Results

Based on the results, the single beam penumbra widths obtained from simulation data for 4, 8, 14 and 18 mm collimator sizes along X axis were 0·75, 0·77, 0·90 and 0·92 mm, respectively. The data for 201 beams obtained from simulation were 2·61, 4·80, 7·92 and 9·81 mm along X axis and 1·31, 1·60, 1·91 and 2·14 mm along Z axis and from film dosimetry were 3·21, 4·90, 8·00 and 10·61 mm along X axis and 1·22, 1·69, 2·01 and 2·25 mm along Z axis, respectively.

Conclusion

The differences between measured and simulated penumbra widths are in an acceptable range. However, for more precise measurement in the penumbra region in which dose gradient is high, Monte Carlo simulation is recommended.

Copyright
Corresponding author
Author for correspondence: Ghazale Geraily, Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Iran. Email: gh-geraily@sina.tums.ac.ir
References
Hide All
1. Nabavi, M, Nedaie, H A, Salehi, N, Naderi, M. Stereotactic radiosurgery/radiotherapy: a historical review.. IJMP 2014; 11 (1): 156167.
2. Tyler, M, Liu, P Z, Chan, K W et al. Characterization of small-field stereotactic radiosurgery beams with modern detectors. Phys Med Biol 2013; 58 (21): 75957608.
3. Xiong, W, Huang, D, Lee, L et al editors. Implementation of Monte Carlo simulations for the Gamma Knife system. Journal of Physics: Conference Series; 2007: IOP Publishing.
4. Al-Dweri, F M, Lallena, A M. A simplified model of the source channel of the Leksell Gamma Knife®: testing multisource configurations with PENELOPE. Phys Med Biol 2004; 49 (15): 34413453.
5. Moskvin, V, DesRosiers, C, Papiez, L, Timmerman, R, Randall, M, DesRosiers, P. Monte Carlo simulation of the Leksell Gamma Knife®: I. Source modelling and calculations in homogeneous media. Phys Med Biol 2002; 47 (12): 1995.
6. Day, M, Lambert, G, Locks, S. The effect of secondary electron spread on the penumbra in high energy photon beam therapy. Br J Radiol 1990; 63 (748): 278285.
7. Yarahmadi, M, Allahverdi, M, Nedaie, H A, Asnaashari, K, Vaezzadeh, S A, Sauer, O A. Improvement of the penumbra for small radiosurgical fields using flattening filter free low megavoltage beams. Zeitschrift für Medizinische Physik 2013; 23 (4): 291299.
8. O’Malley, L, Pignol, J-P, Beachey, D J, Keller, B M, Presutti, J, Sharpe, M. Improvement of radiological penumbra using intermediate energy photons (IEP) for stereotactic radiosurgery. Phys Med Biol 2006; 51 (10): 2537.
9. Laughlin, J S, Mohan, R, Kutcher, G J. Choice of optimum megavoltage for accelerators for photon beam treatment. Int J Radiat Oncol Biol Phys 1986; 12 (9): 15511557.
10. Oh, S A, Kang, M K, Yea, J W, Kim, S K, Oh, Y K. Study of the penumbra for high-energy photon beams with Gafchromic™ EBT2 films. J Korean Phy Soc 2012; 60 (11): 19731976.
11. Romano, F, Sabini, M, Cuttone, G, Russo, G, Mongelli, V, Foroni, R. editors. Geant4-based Monte Carlo Simulation of the Leksell Gamma Knife®. Nuclear Science Symposium Conference Record, 2007 NSS'07 IEEE; 2007: IEEE.
12. Cheung, J Y, Yu, K, Ho, R T, Yu, C. Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit. Med Phys 1999; 26 (7): 12521256.
13. Elekta. Leksell Gamma Knife C, Software upgrade, Instruction for use 2004.
14. Najafi, M, Geraily, G, Shirazi, A, Esfahani, M, Teimouri, J. Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit. Med Dosim 2017; 42 (3): 159168.
15. Rogers, D, Walters, B, Kawrakow, I. BEAMnrc users manual. Nrc Report Pirs 2009; 509: 12.
16. Guerrero, M, Li, X A, Ma, L. A technique to sharpen the beam penumbra for Gamma Knife radiosurgery. Phys Med Biol 2003; 48 (12): 1843.
17. Keller, B M, Beachey, D J, Pignol, J P. Experimental measurement of radiological penumbra associated with intermediate energy x‐rays and small radiosurgery field sizes. Med Phys 2007; 34 (10): 39964002.
18. Keller, B M, Pignol, J P, Presutti, J, Beachey, D J. Intermediate energy photons to improve dose gradient, conformality, and homogeneity: Potential benefits for small field intracranial radiosurgery. Med Phys 2009; 36 (1): 3339.
19. Bilge, H, Ozen, Z, Senkesen, O, Kucucuk, H, Cakir, A, Sengoz, M. Determination of output factors for the Leksell gamma knife using ion chamber, thermoluminescence detectors and films. J BUON 2006; 11 (2): 223.
20. Bertolino, N. Experimental validation of Monte Carlo simulation of Leksell Gamma Knife Perfexion stereotactic radiosurgery system. Ms thesis. University of Milan, Milan, Italy. 2008-2009.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Journal of Radiotherapy in Practice
  • ISSN: 1460-3969
  • EISSN: 1467-1131
  • URL: /core/journals/journal-of-radiotherapy-in-practice
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Keywords

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed