Practical lookup tables for ensuring target coverage in a clinical setup for skin cancer electron therapy

  • Yongsook C. Lee (a1) and Yongbok Kim (a1)

Abstract

Aim

To create practical lookup tables containing percent depth dose (PDD) and profile parameters of electron beams and to demonstrate clinical application of the lookup tables to skin cancer treatment to ensure target coverage in a clinical setup.

Materials and methods

For 6 and 9 MeV electron energies, PDDs and profiles at clinically relevant depths [i.e., R95 (distal depth of 95% maximum dose), R90, R85 and R80] were measured in water at 100 cm source-to-surface distance for an 10×10 cm2 open field and circular cutouts with diameters of 4, 5, 6, 7 and 8 cm. Then PDD parameters along with profile parameters such as width of isodose lines and penumbra at the clinically relevant depths were determined. Output factors for the cutouts were measured at dmax in water and solid water.

Results

With PDD and profile parameters, dosimetry lookup tables were generated. Based upon the lookup tables, target coverage at prescribed depths was retrospectively reviewed for three skin cancer cases. The lookup tables suggested larger cutouts for adequate target coverage.

Findings

Dosimetry lookup tables for electron beam therapy should include profile parameters at clinically relevant depths and be provided to clinicians to ensure target coverage in a clinical setup.

Copyright

Corresponding author

Correspondence to: Yongbok Kim, Department of Radiation Oncology, The University of Arizona, 1501N Campbell Avenue, P.O. Box 245081, Tucson, AZ 85724-5081, USA. Tel: +1 520 694 7427. Fax: +1 520 694 0228. E-mail: yongbokkim@email.arizona.edu

References

Hide All
1. Das, I, Cheng, C W, Healey, G A. Optimum field size and choice of isodose lines in electron beam treatment. Int J Radiat Oncol Biol Phys 1995; 31: 157163.
2. Gerbi, B J, Antolak, J A, Deibel, F C et al. Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25. Med Phys 2009; 36: 32393279.
3. Ding, G X, Duggan, D M, Coffey, C W, Shokrani, P, Cygler, J E. First macro Monte Carlo based commercial dose calculation module for electron beam treatment planning – new issues for clinical consideration. Phys Med Biol 2006; 51: 27812799.
4. Popple, R A, Weinber, R, Antolak, J A et al. Comprehensive evaluation of a commercial macro Monte Carlo electron dose calculation implementation using a standard verification data set. Med Phys 2006; 33: 15401551.
5. Xu, Z, Walsh, S E, Telivala, T P, Meek, A G, Yang, G. Evaluation of the eclipse electron Monte Carlo dose calculation for small fields. J Appl Clin Med Phys 2009; 10: 2834.
6. Khan, F M, Doppke, K P, Hogstrom, K R et al. Clinical electron-beam dosimetry: report of AAPM Radiation Therapy Committee Task Group No. 25. Med Phys 1991; 18: 73109.
7. Demir, B, Okutan, M, Cakir, A, Gokel, E, Bilge, H. The effect of oblique electron beams to the surface dose under the bolus. Med Dosim 2009; 34: 311316.
8. Gunhan, B, Kemikler, G, Koca, A. Determination of surface dose and the effect of bolus to surface dose in electron beams. Med Dosim 2003; 28: 193198.
9. Arunkumar, T, Supe, S S, Ravikumar, M, Sathiyan, S, Ganesh, M. Electron beam characteristics at extended source-to-surface distances for irregular cut-outs. J Med Phys 2010; 35: 207214.
10. Washington, C M, Leaver, D T. Principles and Practice of Radiation Therapy, 4th edition. St. Louis, MO: Elsevier Mosby, 2015; 515.
11. Podgorsak, E B. Review of Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy Agency, 2005; 287.
12. Zhang, S, Liengsawangwong, P, Lindsay, P et al. Clinical implementation of electron energy changes of Varian linear accelerators. J Appl Clin Med Phys 2009; 10: 2978.
13. Cygler, J, Li, X A, Ding, G X, Lawrence, E. Practical approach to electron beam dosimetry at extended SSD. Phys Med Biol 1997; 42: 15051514.
14. Khan, F M, Higgins, P D, Gerbi, B J, Deibel, F C, Sethi, A, Mihailidis, D N. Calculation of depth dose and dose per monitor unit for irregularly shaped electron fields. Phys Med Biol 1998; 43: 27412754.

Keywords

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed