Hostname: page-component-76fb5796d-22dnz Total loading time: 0 Render date: 2024-04-26T00:53:08.884Z Has data issue: false hasContentIssue false

AFM Methodology for the Measurement of Silicon Wafer Microroughness

Published online by Cambridge University Press:  15 February 2011

A.G. Gilicinski
Affiliation:
Air Products and Chemicals, Inc., 7201 Hamilton Blvd., Allentown, PA 18195
S.E. Beck
Affiliation:
Air Products and Chemicals, Inc., 7201 Hamilton Blvd., Allentown, PA 18195
R.M. Rynders
Affiliation:
Air Products and Chemicals, Inc., 7201 Hamilton Blvd., Allentown, PA 18195
D.A. Moniot
Affiliation:
Air Products and Chemicals, Inc., 7201 Hamilton Blvd., Allentown, PA 18195
Get access

Abstract

Despite the growing use of atomic force microscopy (AFM) for the measurement of silicon wafer microroughness, no generally accepted method has been developed to deal with issues around accuracy and reproducibility. We review problems that affect these AFM studies and demonstrate the effect of probe tip size on AFM microroughness data. Without knowledge of AFM probe tip geometry, it is impossible to quantitatively compare Ra or RMS microroughness data between different measurements. An experimental solution is to characterize tip sizes during imaging and compare data taken with similar size tips. While this will significantly improve quantitation, it is restrictive in that data taken with different size tips cannot be easily compared. We propose a solution to this problem in the use of power spectral density (PSD) to evaluate microroughness with a “cutoff frequency” at the lateral wavelength where tip effects begin to affect the accuracy of the microroughness measurement. An example of this approach is described

Type
Research Article
Copyright
Copyright © Materials Research Society 1997

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Hartmann, E., Hahn, P. O., Behm, R. J., J. Appl. Phys. 69 (1991) 4273.Google Scholar
2. Ohmi, T., Miyashita, M., Itano, M., Imaoka, T., and Kawanabe, I., IEEE Trans. Electron Dev., 39 (1992) 537.Google Scholar
3. Grundner, M., Graf, D., Hahn, P. O., Schnegg, A., Solid State Technol. 2 (1991) 69.Google Scholar
4. Griffith, J. E., Grigg, D. A., J. Appl. Phys. 74 (1993) R83.Google Scholar
5. Samitsu, Y., Nanotechnology 4 (1993) 236.Google Scholar
6. Gilicinski, A. G., Rynders, R. M., Beck, S. E., Strausser, Y. E., Stets, J. R., Felker, B. S., Bohling, D. A., Diagnostic Techniques in Semiconductor Manufacturing Processes Mater. Res. Soc. Symp. Proc. 324 (1993) p. 391.Google Scholar
7. Beck, S. E. and Gilicinski, A. G., Proceedings of the IEEE/SEMI Advanced Semiconductor Manufacturing Conference (1995) 68.Google Scholar
8. Bullis, W. M., Semiconductor Silicon 1994 Proc.-Electrochem. Soc. 94–10 (1994) 1156.Google Scholar