ECG artifacts and heart period variability: Don't miss a beat!

  • GARY G. BERNTSON (a1) and JEFFREY R. STOWELL (a1)
    • Published online by Cambridge University Press: 23 July 2016

Abstract

The impact of artifacts on estimates of heart period variability were evaluated by modeling the effects of missed R-waves and spurious R-wave detections in actual and simulated heart period series. Results revealed that even a single artifact, occurring within a 128-s interbeat interval series, can impart substantial spurious variance into all commonly analyzed frequency bands, including that associated with respiratory sinus arrhythmia. In fact, the spurious variance introduced by a single artifact may be greater than that associated with true basal heart period variability and can far exceed typical effect sizes in psychophysiological studies. The effects of artifacts are not related to a specific analytical method and are apparent in both frequency and time domain analyses. Results emphasize the importance of artifact detection and resolution for studies of heart period variability.

Copyright

Corresponding author

Address reprint requests to: Gary G. Berntson, The Ohio State University, 48 Townshend Hall, Columbus, OH 43210, USA. E-mail: berntson.2@osu.edu.
Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Psychophysiology
  • ISSN: 0048-5772
  • EISSN: 1469-8986
  • URL: /core/journals/psychophysiology
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Keywords

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed