Hostname: page-component-76d6cb85b7-xh428 Total loading time: 0 Render date: 2026-07-17T16:00:13.175Z Has data issue: false hasContentIssue false

Wobbling nature of gamma passing rate as a function of calibration field sizes in patient-specific quality assurance

Published online by Cambridge University Press:  11 January 2024

Sathiya Raj*
Affiliation:
Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Dr. M.H. Marigowda Road, Bengaluru, India
Venugopal Sundaram
Affiliation:
Meherbai Tata Memorial Hospital, R53J+3HJ, Stocking Road, Northern Town, Bistupur, Jamshedpur, Jharkhand, India
Henry Finlay Godson
Affiliation:
Department of Radiation Oncology, Christian Medical College Vellore, Ranipet Campus, Kilminnal Village, Ranipet District, TN, India
Retna John
Affiliation:
Department of Radiation Oncology, Christian Medical College Vellore, Ranipet Campus, Kilminnal Village, Ranipet District, TN, India
*
Corresponding author: Sathiya Raj; Email: sathiarajmedphy@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose:

This study aimed to investigate the influence of calibration field size on the gamma passing rate (GPR) in patient-specific quality assurance (PSQA).

Methods:

Two independent detectors, PTW OCTAVIUS 4D (4DOCT) and Arc Check, were utilised in volumetric modulated arc therapy plans for 26 patients (14 with Arc Check and 12 with 4DOCT). Plans were administered using Varian Unique machine (with 4DOCT) and Varian TrueBeam (with Arc Check), each employing different calibration factors (CFs): 4 × 4, 6 × 6, 8 × 8, 10 × 10, 12 × 12 and 15 × 15 cm2 field sizes. Gamma analysis was conducted with 2%2mm, 2%3mm and 3%3mm gamma criteria.

Results:

GPR exhibited variations across different CFs. GPR demonstrated an increasing trend below 10 × 10 cm² CFs, while it displayed a decreasing trend above 10 × 10 cm². Both detectors exhibited similar GPR patterns. The correlation between 4DOCT and Arc Check was strong in tighter criteria (2%2mm) with an R² value of 0·9957, moderate criteria (2%3mm) with an R² value of 0·9868, but reduced in liberal criteria (3%3mm) with an R² value of 0·4226.

Conclusion:

This study demonstrates that calibration field sizes significantly influence GPR in PSQA. This study recommends the plan specific calibration field must obtain to calibrate the QA devices for modulated plans.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Simple calibration square field not true representation of the true complex segments of the VMT/IMRT plans

Figure 1

Figure 2. GPR for different calibration field size in 4DOCT

Figure 2

Figure 3. GPR for different calibration field size (Arc Check)

Figure 3

Figure 4. (a), (b), (c): Correlation of GPR between 4DOCT and Arc Check for (a = 2%2mm, b = 2%3mm and c = 3%3mm)

Figure 4

Table 1. Uncertainty budget

Figure 5

Table 2. Measured and TPS dose comparison as function of calibration field size in 4DOCT

Figure 6

Table 3. Measured and TPS dose comparison as function of calibration field size in Arc Check

Supplementary material: PDF

Raj et al. supplementary material

Raj et al. supplementary material

Download Raj et al. supplementary material(PDF)
PDF 1 MB