Head mirror versus headlight: illumination, visual identification and visual acuity for otolaryngological examination

Published online by Cambridge University Press:  09 July 2013

C-H Lin
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
H-T Hsu
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
P-Y Chen
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Shuang Ho Hospital, Taipei, Taiwan
L K Huon
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Cathay General Hospital, Taipei, Taiwan
Y-Z Lin
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
S-H Hung*
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
*
Address for correspondence: Dr Shih-Han Hung, Department of Otolaryngology, School of Medicine, Taipei Medical University, 250 Wusing St, Sinyi District, Taipei City 110, Taiwan Fax: +886 2 27359903 E-mail: seedturtle@gmail.com

Abstract

Objective:

To investigate and compare the performance of head mirrors and headlights during otolaryngological examination.

Methods:

The illuminance and illumination field of each device were measured and compared. Visual identification and visual acuity were also measured, in 13 medical students and 10 otolaryngology specialists.

Results:

The illuminance (mean ± standard deviation) of the LumiView, Kimscope 1 W and Kimscope 3 W headlights and a standard head mirror were 352.3 ± 9, 92.3 ± 4.5, 438 ± 15.7 and 68.3 ± 1.2 lux, respectively. The illumination field of the head mirror (mean ± standard deviation) was 348 ± 29.8 grids, significantly greater than that of the Kimscope 3 W headlight (183 ± 9.2 grids) (p = 0.0017). The student group showed no statistically significant difference between visual identification with the best headlight and the head mirror (score means ± standard deviations: 56.2 ± 9 and 53.3 ± 14.1, respectively; p = 0.3). The expert group scored significantly higher for visual identification with head mirrors versus headlights (59.7 ± 3.3 vs 55.2 ± 5.8, respectively; p = 0.0035), but showed no difference for visual acuity.

Conclusion:

Despite the advantages of headlight illumination, head mirrors provided better, shadow-free illumination. Despite no differences amongst students, head mirrors performed better than headlights in experienced hands.