Hostname: page-component-6766d58669-6mz5d Total loading time: 0 Render date: 2026-05-17T15:51:24.504Z Has data issue: false hasContentIssue false

Reduction of airborne and surface-borne bacteria in a medical center burn intensive care unit using active, upper-room, germicidal ultraviolet (GUV) disinfection

Published online by Cambridge University Press:  25 October 2023

Linda D. Lee*
Affiliation:
UV Angel, Grand Haven, Michigan
Louise Lie
Affiliation:
Loyola University Medical Center, Maywood, Illinois
Michael Bauer
Affiliation:
Loyola University Medical Center, Maywood, Illinois
Berenice Bolanos
Affiliation:
Loyola University Medical Center, Maywood, Illinois
Russell N. Olmsted
Affiliation:
Trinity Health, Livonia, Michigan
Jay K. Varma
Affiliation:
Weill Cornell Medical College, New York, New York
Jorge P. Parada
Affiliation:
Loyola University Medical Center, Maywood, Illinois Loyola University Chicago, Chicago, Illinois
*
Author for correspondence: Linda D. Lee, DrPH, MBA, UV Angel, 233 Washington Ave, Suite L1, Grand Haven, MI 49417. E-mail: llee@uvangel.com
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To determine the effectiveness of active, upper-room, germicidal ultraviolet (GUV) devices in reducing bacterial contamination in patient rooms in air and on surfaces as a supplement to the central heating, ventilation, and air conditioning (HVAC) air handling unit (AHU) with MERV 14 filters and UV-C disinfection.

Methods:

This study was conducted in an academic medical center, burn intensive care unit (BICU), for 4 months in 2022. Room occupancy was monitored and recorded. In total, 402 preinstallation and postinstallation bacterial air and non–high-touch surface samples were obtained from 10 BICU patient rooms. Airborne particle counts were measured in the rooms, and bacterial air samples were obtained from the patient-room supply air vents and outdoor air, before and after the intervention. After preintervention samples were obtained, an active, upper-room, GUV air disinfection system was deployed in each of the patient rooms in the BICU.

Results:

The average levels of airborne bacteria of 395 CFU/m3 before GUV device installation and 37 CFU/m3 after installation indicated an 89% overall decrease (P < .0001). Levels of surface-borne bacteria were associated with a 69% decrease (P < .0001) after GUV device installation. Outdoor levels of airborne bacteria averaged 341 CFU/m3 in March before installation and 676 CFU/m3 in June after installation, but this increase was not significant (P = .517).

Conclusions:

Significant reductions in air and surface contamination occurred in all rooms and areas and were not associated with variations in outdoor air concentrations of bacteria. The significant decrease of surface bacteria is an unexpected benefit associated with in-room GUV air disinfection, which can potentially reduce overall bioburden.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2023. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Table 1. Room Volume, Airflow, and Occupancy

Figure 1

Figure 1. Comparison of before versus after overall room average colony-forming units (CFU) for germicidal ultraviolet light (GUV) in air (CFU/m3) and on surfaces (CFU/plate).

Figure 2

Table 2. Average Mitigated CFUs for Airborne and Surface Samples With Aspin-Welch Unequal Variance T-Test P Values Before and After Installation of a UV-C Air Cleaning Unit

Figure 3

Figure 2. Comparison of total airborne colony-forming units (CFU) in each room with and without germicidal ultraviolet light (GUV) device. Error bars show the standard error of the mean (SEM) values.

Figure 4

Figure 3. Comparison of total surface-borne colony-forming units (CFU) in each room with and without a germicidal ultraviolet light (GUV) device. Error bars show the SEM values.

Figure 5

Table 3. Average Supply Hood CFUs for Airborne Samples With Aspin-Welch Unequal Variance T Test P Values Before and After Installation of a UV-C Air Cleaning Unit

Figure 6

Table 4. Particle Counts in Rooms and Supply Hoods Before and After Installation of a UV-C Air Cleaning Unit