Hostname: page-component-89b8bd64d-shngb Total loading time: 0 Render date: 2026-05-07T07:19:08.837Z Has data issue: false hasContentIssue false

Radiation dose diode measurements for in vivo dosimetry of IMRT plans: accuracy and technical considerations

Published online by Cambridge University Press:  06 February 2025

Ali H. Abdelkader
Affiliation:
Department of Radiation Oncology, American University of Beirut Medical Center, Beirut, Lebanon
Ali M. AlZein
Affiliation:
Department of Radiation Oncology, American University of Beirut Medical Center, Beirut, Lebanon
Bilal H. Shahine*
Affiliation:
Department of Radiation Oncology, American University of Beirut Medical Center, Beirut, Lebanon
*
Corresponding author: Bilal Shahine; Email: bs39@aub.edu.lb
Rights & Permissions [Opens in a new window]

Abstract

Introduction:

In vivo diode dosimetry proved its efficacy as a patient-specific quality assurance tool for 3DCRT. Its usefulness in IMRT is not sufficiently investigated.

Methods:

Four step-and-shoot IMRT plans were generated for different treatment sites using an Alderson Rando anthropomorphic phantom. Two in vivo semiconductor diodes were used to perform 33 entrance radiation dose measurements at central axis and max beam point dose on a water-equivalent slabs based on the anthropomorphic plans. Measured values were compared to planned ones, and detailed analysis per segment of each IMRT field was made to infer the reason behind the discrepancies of measurements from expectations.

Results:

Point dose measurements were performed on a beam and a segment basis. Agreement within ±5% action level with planned dose was 27% of beams at central axis versus 45% at max point dose and 10% of segments at central axis versus 25% at max point dose. For ±10% action level, 70% of beams at central axis versus 73% at max point dose, and 15% of segments at central axis versus 41% at max point dose. Classification of segments resulted in a significance of .021 for measuring at positions unaffected by the MLC partial and total blockage.

Conclusion:

Diode measurements are recommended at maximum dose coordinates for open beams/segments for more accurate patient dose verification results as part of in vivo dosimetry. This is important for limited resources centres treating with sMLC IMRT.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press
Figure 0

Figure 1. The axial plane of the anthropomorphic phantom scan for the four treatment plans: (a) brain, (b) left breast, (c) left lung, and (d) prostate.

Figure 1

Figure 2. MLCs totally and partially blocking the max dose and central axis measurement points in segments 5 and 7 of beams (a) ‘RPO’ and (b) ‘RPO2’ in the left lung and prostate cases, respectively. (c) The central axis and max dose positions at which diodes A and B are placed, respectively, in the ‘LAO’ beam of the prostate plan.

Figure 2

Table 1. The expected and measured doses along with their corresponding percentage differences at both positions for all prostate beams and segments of the LPO beam

Figure 3

Table 2. Percentage of beams and segments having measured dose within ±5% and ±10% of the expected one for the different treatment plans at the MDP and CA positions

Figure 4

Table 3. (a) Percentage of totally blocked (TB) & partially blocked (PB) and open (O) segments according to the measurement points with respect to the MLCs for each plan and (b) the percentage of PB & O segments having the measured dose within ±5% and ±10% of the expected one at both measurement positions

Figure 5

Table 4. Percentage of segments having open measurement points and the beams/segments MAE at the two examined positions for the four treatment plans