One layer at a time: the use of 3D printing in the fabrication of cadmium-free electron field shaping devices
- Michael J. Moore, Ronald Snelgrove, Johnson Darko, Ernest K. Osei
-
- Journal:
- Journal of Radiotherapy in Practice / Volume 21 / Issue 2 / June 2022
- Published online by Cambridge University Press:
- 14 December 2020, pp. 222-227
-
- Article
- Export citation
-
Introduction:
Electron blocks are typically composed of a low melting point alloy (LMPA), which is poured into an insert frame containing a manually placed Styrofoam aperture negative used to define the desired field shape. Current implementations of the block fabrication process involve numerous steps which are subjective and prone to user error. Occasionally, bowing of the sides of the insert frame is observed, resulting in premature frame decommissioning. Recent works have investigated the feasibility of utilising 3D printing technology to replace the conventional electron block fabrication workflow; however, these approaches involved long print times, were not compatible with commonly used cadmium-free LMPAs, and did not address the problem of insert frame bowing. In this work, we sought to develop a new 3D printing technique that would remedy these issues.
Materials and Methods:Electron cutout negatives and alignment jigs were printed using Acrylonitrile Butadiene Styrene, which does not warp at the high temperatures associated with molten cadmium-free alloys. The accuracy of the field shape produced by electron blocks fabricated using the 3D printed negatives was assessed using Gafchromic film and beam profiler measurements. As a proof-of-concept, electron blocks with off-axis apertures, as well as complex multi-aperture blocks to be used for passive electron beam intensity modulation, were also created.
Results:Film and profiler measurements of field size were in excellent agreement with the values calculated using the Eclipse treatment planning system, showing less than a 1% difference in line profile full-width at half-maximum. The multi-aperture electron blocks produced fields with intensity modulation ≤3.2% of the theoretically predicted value. Use of the 3D printed alignment jig – which has contours designed to match those of the insert frame – was found to reduce the amount of frame bowing by factors of 1.8 and 2.1 in the lateral and superior–inferior directions, respectively.
Conclusions:The 3D printed ABS negatives generated with our technique maintain their spatial accuracy even at the higher temperatures associated with cadmium-free LMPA. The negatives typically take between 1 and 2 hours to print and have a material cost of approximately $2 per patient.
Dosimetric evaluation of SBRT treatment plans of non-central lung tumours: clinical experience
- Ernest Osei, Johnson Darko, Steph Swanson, Katrina Fleming, Ronald Snelgrove, Anmol Bhangu, Darin Gopaul
-
- Journal:
- Journal of Radiotherapy in Practice / Volume 21 / Issue 2 / June 2022
- Published online by Cambridge University Press:
- 09 December 2020, pp. 179-199
-
- Article
- Export citation
-
Objectives:
Lung cancer is the most commonly diagnosed cancer in Canada and the leading cause of cancer-related mortality in both men and women in North America. Surgery is usually the primary treatment option for early-stage non-small cell lung cancer (NSCLC). However, for patients who may not be suitable candidates for surgery, stereotactic body radiation therapy (SBRT) is an alternative method of treatment. SBRT has proven to be an effective technique for treating NSCLC patients by focally administering high radiation dose to the tumour with acceptable risk of toxicity to surrounding healthy tissues. The goal of this comprehensive retrospective dosimetric study is to compare the dosimetric parameters between three-dimensional conformal radiation therapy (3DCRT) and volumetric-modulated arc therapy (VMAT) lung SBRT treatment plans for two prescription doses.
Methods:We retrospectively analysed and compared lung SBRT treatment plans of 263 patients treated with either a 3DCRT non-coplanar or with 2–3 VMAT arcs technique at 48 Gy in 4 fractions (48 Gy/4) or 50 Gy in 5 fractions (50 Gy/5) prescribed to the planning target volume (PTV), typically encompassing the 80% isodose volume. All patients were treated on either a Varian 21EX or TrueBeam linear accelerator using 6-MV or 10-MV photon beams.
Results:The mean PTV V95% and V100% for treatment plans at 48 Gy/4 are 99·4 ± 0·6% and 96·0 ± 1·0%, respectively, for 3DCRT and 99·7 ± 0·4% and 96·4 ± 3·4%, respectively, for VMAT. The corresponding mean PTV V95% and V100% at 50 Gy/5 are 99·0 ± 1·4% and 95·5 ± 2·5% for 3DCRT and 99·5 ± 0·8% and 96·1 ± 1·6% for VMAT. The CIRI and HI5/95 for the PTV at 48 Gy/4 are 1·1 ± 0·1 and 1·2 ± 0·0 for 3DCRT and 1·0 ± 0·1 and 1·2 ± 0·0 for VMAT. The corresponding CIRI and HI5/95 at 50 Gy/5 are 1·1 ± 0·1 and 1·3 ± 0·1 for 3DCRT and 1·0 ± 0·1 and 1·2 ± 0·0 for VMAT. The mean R50% and D2cm at 48 Gy/4 are 5·0 ± 0·8 and 61·2 ± 7·0% for 3DCRT and 4·9 ± 0·8 and 57·8 ± 7·9% for VMAT. The corresponding R50% and D2cm at 50 Gy/5 are 4·7 ± 0·5 and 65·5 ± 9·4% for 3DCRT and 4·7 ± 0·7 and 60·0 ± 7·2% for VMAT.
Conclusion:The use of 3DCRT or VMAT technique for lung SBRT is an efficient and reliable method for achieving dose conformity, rapid dose fall-off and minimising doses to the organs at risk. The VMAT technique resulted in improved dose conformity, rapid dose fall-off from the PTV compared to 3DCRT, although the magnitude may not be clinically significant.