Neuroradiology Using Secure Mobile Device Review

Published online by Cambridge University Press:  05 April 2016

Privia A. Randhawa
Affiliation:
Calgary Stroke Program, Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada Department of Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada
William Morrish
Affiliation:
Calgary Stroke Program, Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada Department of Radiology and Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada
John T. Lysack
Affiliation:
Calgary Stroke Program, Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada Departments of Radiology, Clinical Neurosciences and Surgery, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada
William Hu
Affiliation:
Calgary Stroke Program, Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada Department of Radiology and Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada
Mayank Goyal
Affiliation:
Calgary Stroke Program, Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada Department of Radiology and Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada
Michael D. Hill
Affiliation:
Calgary Stroke Program, Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada Department of Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada The Hotchkiss Brain Institute, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
Rights & Permissions [Opens in a new window]

Abstract

Background: Image review on computer-based workstations has made film-basedreview outdated. Despite advances in technology, the lack of portability ofdigital workstations creates an inherent disadvantage. As such, we sought todetermine if the quality of image review on a handheld device is adequatefor routine clinical use. Methods: Six CT/CTA cases and six MR/MRA cases were independently reviewed bythree neuroradiologists in varying environments: high and low ambient lightusing a handheld device and on a traditional imaging workstation in idealconditions. On first review (using a handheld device in high ambient light),a preliminary diagnosis for each case was made. Upon changes in reviewconditions, neuroradiologists were asked if any additional features wereseen that changed their initial diagnoses. Reviewers were also asked tocomment on overall clinical quality and if the handheld display was ofacceptable quality for image review. Results: After the initial CT review in high ambient light, additionalfindings were reported in 2 of 18 instances on subsequent reviews.Similarly, additional findings were identified in 4 of 18 instances afterthe initial MR review in high ambient lighting. Only one of these sixadditional findings contributed to the diagnosis made on the initialpreliminary review. Conclusions: Use of a handheld device for image review is of adequate diagnosticquality based on image contrast, sharpness of structures, visible artefactsand overall display quality. Although reviewers were comfortable with usingthis technology, a handheld device with a larger screen may bediagnostically superior.

Résumé

Utilisation d’appareils mobiles sécurisés dans le cadre d’analyses neuroradiologiques. Contexte: L’analyse d’images à des postes de travail informatisés a renduobsolète l’impression de ces images. En dépit des progrès de la technologie,la pénurie de tels postes de travail représente assurément un inconvénient.À cet égard, nous avons tenté de déterminer dans quelle mesure la qualitédes images analysées au moyen d’appareils mobiles pouvait convenir à uneutilisation clinique de routine. Méthodes: De manière indépendante, trois neuroradiologistes ont examiné sixcas de patients soumis à la tomodensitométrie (TDM) et à l’angiographie partomodensitométrie ainsi que six autres cas soumis à des examens derésonnance magnétique et d’angiographie par résonance magnétique. Pour cefaire, ils ont utilisé des appareils mobiles alors que la luminositéambiante était élevée ou faible mais aussi, dans des conditions idéales, despostes de travail pour imagerie tout à fait courants. Dans un premier temps,à l’aide d’une luminosité ambiante élevée, ils ont établi pour chacun desdouze cas un diagnostic préliminaire en utilisant un appareil mobile. Aprèsavoir modifié les conditions d’analyse, on a demandé aux neuroradiologistessi quelque autre aspect observé pouvait modifier leur diagnosticpréliminaire. On leur a également demandé de se prononcer sur la qualitéclinique générale de l’analyse menée et sur la qualité d’imagerie fourniepar les appareils mobiles. Résultats: Après avoir analysé les cas de TDM à l’aide d’une luminositéambiante élevée, des analyses ultérieures ont signalé d’autres observationsdans 2 cas sur 18. À la suite de l’analyse initiale d’angiographie partomodensitométrie à l’aide d’une luminosité ambiante élevée, on a pu noter,dans la même veine, des observations additionnelles dans 4 cas sur 18. Deces 6 observations, seulement une a permis d’affiner le diagnostic établilors de l’analyse initiale. Conclusions: L’utilisation d’appareils mobiles pour analyser des images permetde poser de bons diagnostics. Ces derniers reposent sur une bonne qualitéd’affichage ainsi que sur des images bien contrastées, des structures netteset d’éventuels artefacts bien visibles. Bien que les radioneurologistesévaluateurs aient utilisé avec facilité ces appareils, des écrans pluslarges pourraient s’avérer plus efficaces sur le plan diagnostique.

Information

Type
Original Articles
Copyright
Copyright © The Canadian Journal of Neurological Sciences Inc. 2016 
Figure 0

Table 1 Case List

Figure 1

Table 2 Ambient Light Intensity

Figure 2

Table 3 CT Case Review

Figure 3

Table 4 MR Case Review

Figure 4

Table 5 Qualitative Assessment