Hostname: page-component-6766d58669-h8lrw Total loading time: 0 Render date: 2026-05-17T21:14:52.967Z Has data issue: false hasContentIssue false

Optimizing the configuration of plasma radiation detectors in the presence of uncertain instrument response and inadequate physics

Published online by Cambridge University Press:  06 January 2023

P.F. Knapp*
Affiliation:
Sandia National Laboratories, Albuquerque, NM 87185, USA
W.E. Lewis
Affiliation:
Sandia National Laboratories, Albuquerque, NM 87185, USA
V.R. Joseph
Affiliation:
Stewart School of Industrial and Systems Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USA
C.A. Jennings
Affiliation:
Sandia National Laboratories, Albuquerque, NM 87185, USA
M.E. Glinsky
Affiliation:
Sandia National Laboratories, Albuquerque, NM 87185, USA
*
Email address for correspondence: pfknapp@sandia.gov
Rights & Permissions [Opens in a new window]

Abstract

We present a general method for optimizing the configuration of an experimental diagnostic to minimize uncertainty and bias in inferred quantities from experimental data. The method relies on Bayesian inference to sample the posterior using a physical model of the experiment and instrument. The mean squared error (MSE) of posterior samples relative to true values obtained from a high fidelity model (HFM) across multiple configurations is used as the optimization metric. The method is demonstrated on a common problem in dense plasma research, the use of radiation detectors to estimate physical properties of the plasma. We optimize a set of filtered photoconducting diamond detectors to minimize the MSE in the inferred X-ray spectrum, from which we can derive quantities like the electron temperature. In the optimization we self-consistently account for uncertainties in the instrument response with appropriate prior probabilities. We also develop a penalty term, acting as a soft constraint on the optimization, to produce results that avoid negative instrumental effects. We show results of the optimization and compare with two other reference instrument configurations to demonstrate the improvement. The MSE with respect to the total inferred X-ray spectrum is reduced by more than an order of magnitude using our optimized configuration compared with the two reference cases. We also extract multiple other quantities from the inference and compare with the HFM, showing an overall improvement in multiple inferred quantities like the electron temperature, the peak in the X-ray spectrum and the total radiated energy.

Information

Type
Research Article
Copyright
Copyright © The Author(s), 2023. Published by Cambridge University Press
Figure 0

Figure 1. Network describing our experimental inference problem. Model parameters $\theta$ with appropriate prior distributions are fed into the forward physics model $f(\boldsymbol {d}; \theta )$ producing deterministic output quantity $y$. Model output is fed into the diagnostic models whose behaviours are controlled by the configuration choices $z_i$ and stochastic calibration values $\xi _i$. The output of the diagnostic models are compared with experimental observations $O_i$. Additional unobserved quantities of interest are computed as $Y = h(y)$.

Figure 1

Algorithm 1 Instrument Optimization

Figure 2

Figure 2. Pairwise distributions of emissivity-weighted temperature and liner areal density as well as total X-ray output from the ensemble of one-dimensional simulations. (a) Distribution of liner areal density with temperature. (b) Distribution of $\log _{10}$ of the X-ray output with temperature. (c) Distribution of $\log _{10}$ of the X-ray output with liner areal density. Grey points show the entire dataset, blue points show those used for validation and magenta show those used for training.

Figure 3

Figure 3. (a) Best value of the optimization metric as a function of iteration for runs initialized with five different seeds. The run in green arrived at the best solution. (b) The spectral response of each of the five detectors including the filter for the best solution out of the five runs.

Figure 4

Table 1. Final configuration of filter material, filter thickness and detector sensitivity for each of the 5 PCD elements determined through the optimization.

Figure 5

Figure 4. (a) Spectral response for the ‘MagLIF’ reference configuration. (b) Spectral response for the ‘best guess’ reference case.

Figure 6

Figure 5. Plots showing the posterior spectrum for three different cases in the validation set inferred using each of the different instrument configurations. The top row shows the spectra inferred using the standard MagLIF configuration, the middle row using the best guess configuration, and the bottom row the optimum. The dashed black lines show the true spectrum, the solid lines show the median of the posterior, and the shaded bands show the credible intervals.

Figure 7

Figure 6. The difference in the posterior inferred and true spectra normalized to the true spectrum, $\varDelta _\epsilon$ for each of the 16 validation cases considered. The dashed black line shows a value of 0, indicating perfect agreement. The solid blue line shows the mean and the light and dark shaded regions show the $68\,\%$ and $95\,\%$ credible intervals. Curves are artificially offset for clarity.

Figure 8

Table 2. Summary of performance of each of the metrics discussed on the validation dataset. The first column shows the MSE defined on the validation set and the second column shows the peak voltage produced over the validation set with the given configuration. Remaining columns summarize the features computed using (4.2) and (4.3) for the peak intensity, photon energy of the peak and the continuum slope.

Figure 9

Figure 7. Scatter plots showing the inferred vs. true values of the peak intensity of the spectrum (a), the location of the peak in the spectrum (b) and high energy slope of the spectrum, as defined in the text (c). The true value is shown on the abscissa and the inferred value on the ordinate. Median values inferred using the standard MagLIF configuration are shown in blue, with the best guess configuration are shown in orange and the optimum configuration are shown orange. The error bars indicate the 16 %–84 % credible interval.

Figure 10

Figure 8. Scatter plots showing the inferred vs. true values of the temperature (a), liner areal density (b) and X-ray output (c). The true value is shown on the abscissa and the inferred value on the ordinate. Median values inferred using the standard MagLIF configuration are shown in blue, with the best guess configuration are shown in orange and the optimum configuration are shown orange. The error bars indicate the 16 %–84 % credible interval.

Figure 11

Table 3. Summary of the performance of each configuration using (4.2) and (4.3) for the two model parameters temperature and $\rho R_\ell$ as well as the total radiated output.

Figure 12

Figure 9. Corner plot showing the posterior distribution of model parameters for one of the cases in the validation set. The plots on the diagonal show the marginalized posteriors of temperature liner areal density and total output. The off-diagonal plots show the pairwise joint distributions. The posterior obtained using the standard MagLIF configuration, best guess configuration and optimum configuration are shown in blue, orange and green, respectively.

Figure 13

Table 4. Filter materials along with numerical index used in the optimization.

Figure 14

Figure 10. Peak voltages as a function of the hyperparameter $\lambda$. Blue circles show the peak voltage found over the entire validation dataset, orange circles show the mean of the peak voltage found for each case over the dataset.

Figure 15

Figure 11. Peak voltages as a function of the hyperparameter $\alpha$. Blue circles show the peak voltage found over the entire validation dataset, orange circles show the mean of the peak voltage found for each case over the dataset.