Hostname: page-component-8448b6f56d-gtxcr Total loading time: 0 Render date: 2024-04-18T23:05:28.844Z Has data issue: false hasContentIssue false

Non-scanning systems for far-infrared radiation detection from laser-induced plasmas

Published online by Cambridge University Press:  18 May 2015

Chun Li
Affiliation:
Beijing National Laboratory of Condensed Matter Physics, Institute of Physics, CAS, Beijing 100190, PR China
Guo-Qian Liao
Affiliation:
Beijing National Laboratory of Condensed Matter Physics, Institute of Physics, CAS, Beijing 100190, PR China
Yu-Tong Li*
Affiliation:
Beijing National Laboratory of Condensed Matter Physics, Institute of Physics, CAS, Beijing 100190, PR China IFSA Collaborative Innovation Center, Shanghai Jiao Tong University, Shanghai 200240, PR China
*
Correspondence to: Y.-T. Li, Laboratory of Optical Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, PR China. Email: ytli@iphy.ac.cn

Abstract

With the development of ultraintense terawatt (TW) and petawatt (PW) laser systems, powerful terahertz (THz) radiation from laser–plasma interactions has been reported. Plasma-based THz systems, which are usually operated at extremely low repetition rates, call for single-shot diagnostics. In this paper, various state-of-the-art single-shot detection methods are introduced or designed for measurements and applications involved in high-power plasma-based THz sciences.

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
The online version of this article is published within an Open Access environment subject to the conditions of the Creative Commons Attribution licence .
Copyright
© The Author(s) 2015

1. General review on THz detectors

Nowadays, detection of radiation in the terahertz (THz) frequency domain is of great interest in active research[Reference Ferguson and Zhang1, Reference Siegel2]. Generally speaking, the methods involved can be mainly categorized into two types: incoherent (direct) detection and coherent detection. In order to meet the requirements of different applications, designs for specific THz radiation detection systems need to take the following three factors into account: (1) whether the system has sufficiently high sensitivity to allow signals to be extracted from thermal background radiation; (2) whether the data sampling rate allows real-time measurements of ultrafast processes; and (3) whether full characterization of the THz radiation, including amplitude and phase information, can be provided with no or little distortion. The last requirement cannot be achieved other than by using coherent detectors/methods.

Figure 1. Schematic of (a) spectral encoding and (b) spatial encoding scheme.

Spectral information is always considered essential in real applications. For applications that require very high spectral resolution within a known frequency region, heterodyne detection systems based on frequency mixing are preferred. At room temperature, a semiconductor-based detector, such as a planar Schottky-diode mixer, is combined with a local oscillator for frequency downconversion[Reference Tzortzakis, Méchain, Patalano, André, Prade, Franco, Mysyrowicz, Munier, Gheudin, Beaudin and Encrenaz3, Reference Gaidis, Pickett, Smith, Smith, Martin and Siegel4]. The created downshifted signal – the intermediate-frequency (IF) signal – is then filtered and amplified to obtain high sensitivity of detection, the noise equivalent power (NEP) of which is typically on the order of $10^{-19}~\text{W}/\surd \text{Hz}$. Cryogenic-cooling techniques are used in heterodyne systems to provide even smaller NEPs of $10^{-21}{-}10^{-20}~\text{W}/\surd \text{Hz}$. The most widely used superconducting heterodyne detector includes the superconductor–insulator–superconductor (SIS) tunnel junction mixer[Reference Dolan, Phillips and Woody5]. In principle, the detectable frequency range of a heterodyne system is determined by the operating frequency of the local oscillator, which is normally in the sub-THz or low-THz frequency region. The corresponding response time is on the level of picoseconds.

For broadband detection, direct detectors based on thermal absorption are widely used. Room-temperature thermal-type detectors include pyroelectric detectors and Golay cells. Bolometers can also be operated at room temperatures, though helium-cooling technologies are always provided to reduce background thermal noise. Compared with heterodyne devices, direct detectors have typical response times of milliseconds and suffer from a much higher NEP value (typ. $10^{-10}~\text{W}/\surd \text{Hz}$ for room temperature; $10^{-12}{-}10^{-15}~\text{W}/\surd \text{Hz}$ for helium-cooled systems). The responsivity of a thermal-type direct detector usually shows little dependence on radiation wavelength; hence bandpass filters or interferometric measurements are generally involved in providing spectral information of the measured signals. A Fourier transform infrared (FTIR) spectrometer, which is commonly used for the identification and analysis of materials with frequency structures in the THz domain, is operated in conjunction with a far-infrared (FIR) optical interferometer and a direct detector. Moreover, since FIR interferometry is based on the autocorrelation effects of THz pulses, it can also be used to measure the pulse duration of FIR signals[Reference Hamster, Sullivan, Gordon, White and Falcone6].

With the rapid development of ultrafast laser systems, pulsed detectors based on photoconductive or free-space electro-optic (EO) sampling are generally used for coherent detection of broadband THz radiation. The detection scheme excludes the majority of background noise from the time measurement window, and the signal–noise ratio (SNR) of detection surpasses 104 using lock-in technologies[Reference Cai, Brener, Lopata, Wynn, Pfeiffer, Stark, Wu, Zhang and Federici7]. Comparing photoconductive- and EO-based detection systems, the spectral response of the former is affected by the carrier lifetime as well as by antenna structures[Reference Park, Melloch and Weiner8], while the latter, as a pure optical technique, is considered more reliable in providing a flat frequency response over a wider frequency domain by using short probe pulses. Detectable frequencies in excess of 20 and 100 THz have been demonstrated using photoconductive-[Reference Kono, Tani, Gu and Sakai9] and EO-sampling[Reference Kübler, Huber and Leitenstorfer10] detectors, respectively. THz time–domain spectroscopy (THz–TDS), which has been developed on the basis of pulsed THz techniques, is now widely used to study material properties and transient processes in the FIR region.

The effective spectral response range of EO detection systems is ultimately determined by[Reference Gallot and Grischkowsky11]: (1) the group-velocity mismatch (GVM) between THz and probe pulses; and (2) the phonon absorption of EO crystals. ZnTe and GaP are two of the most commonly used sensors for EO sampling. The former shows excellent detection sensitivity around 2 THz due to the small GVM of optical/THz pulses within the tuning range of Ti:sapphire lasers[Reference Wu and Zhang12], while the latter has a high fundamental transverse-optical (TO) phonon resonance frequency near 11 THz[Reference Wu and Zhang13]. Generally speaking, thinner EO crystals tend to provide a wider spectral response at the cost of detection sensitivity. The effective use of a $\langle 110\rangle$-cut ZnTe crystal is narrowed by about 40% (from 4 to 2.5 THz) as the crystal thickness is changed from 0.2 to 1 mm[Reference van Tilborg, Schroeder, Filip, Tóth, Geddes, Fubiani, Huber, Kaindl, Esarey and Leemans14, Reference Nahata, Weling and Heinz15]. In considering the trade-off between high detection sensitivity and broadband spectral response, the type and thickness of EO crystals should be chosen carefully for specific research.

2. Non-scanning schemes based on EO sampling

One of the main drawbacks of conventional EO-sampling techniques is the low acquisition rate. Normally, the read-out time for a single THz waveform would be at least seconds using serial acquisition schemes, which limits the application of this technique for real-time measurements of fast-moving objects or for real-time imaging. Moreover, how to maintain a high SNR of detection while operating the lock-in amplifier with a low time constant has always been an issue in fast scanning[Reference Cai, Brener, Lopata, Wynn, Pfeiffer, Stark, Wu, Zhang and Federici7].

To increase the data acquisition rate, conventional single-channel delay-scanning detection should be replaced by parallel data collection and multi-channel detector arrays. Much effort has been made in developing ‘single-shot’ (non-scanning) EO detection systems since 1998, and in general the techniques are progressing in two directions: spectral encoding[Reference Jiang and Zhang16] and spatial encoding[Reference Shan, Weling, Knoesel, Bartels, Bonn, Nahata, Reider and Heinz17]; see Figures 1(a) and 1(b), respectively. The main idea is to record the entire temporal profile of a THz pulse onto the spectral or spatial profile of a probe pulse for parallel data acquisition by a spectrometer or a charge-coupled device (CCD) camera. However, both schemes have obvious shortcomings. The spectral encoding method, which relies on measurement of the spectral modulation of a chirped probe pulse, has a fundamental temporal resolution limit related to the chirped pulse duration[Reference Sun, Jiang and Zhang18]. The spatial encoding method, which is based on the non-collinear geometry of a collimated short probe pulse along with a collimated THz pulse, retains an excellent time resolution but suffers from a much weaker sensitivity of detection. In addition, the spatial encoding method relies only on the energy/pointing stability of the laser system, while the spectral encoding scheme is also sensitive to spectral instability of the laser.

To achieve both high sensitivity and temporal resolution, a third ‘single-shot’ EO-cased technique has been introduced as a cross-correlation technique by second-harmonic generation (SHG)[Reference Jamison, Shen, MacLeod, Gillespie and Jarpszynski19], as depicted in Figure 2, which can be considered as a combination of the spectral and spatial encoding techniques. However, since a second probe pulse is necessary, the complexity of the setup limits the applicability of this technique in real applications. In order to improve the temporal resolution while retaining the simplicity of measurements included in the spectral encoding technique, a different retrieval algorithm has been applied to the experimental setup depicted in Figure 1(a). The corresponding technique is referred to as ‘single-shot’ in-line spectral interferometry[Reference Kim, Yellampalle, Rodriguez, Averitt, Taylor and Glownia20, Reference Yellampalle, Kim, Rodriguez, Glownia and Taylor21], by which the THz field will be derived from the measured optical spectral modulation through a matrix inversion. However, the algorithm cannot be applied in experiments which involve substantial noise.

Figure 2. Schematic of ‘single-shot’ cross-correlation scheme.

Based on the schemes (including algorithms) shown in Figures 1 and 2, proof-of-principle experiments have been conducted on quasi-continuous-wave lasers[Reference Jiang and Zhang16, Reference Shan, Weling, Knoesel, Bartels, Bonn, Nahata, Reider and Heinz17, Reference Jamison, Shen, MacLeod, Gillespie and Jarpszynski19, Reference Kim, Yellampalle, Rodriguez, Averitt, Taylor and Glownia20] (operated at kHz or MHz). Since at least two laser shots are needed to identify the modulation caused by the THz field, the accuracy of signal recovery depends strongly on the stability of the laser system. Sequential acquisition of the signal and reference pulses will prevent the use of such techniques in high-energy laser systems, which usually operate at low repetition rates and are accordingly less stable.

3. True single-shot THz diagnostics in intense laser–plasma interactions

The development of THz sciences is in great need of high-power THz sources. With the rapid development of ultrashort laser systems, laser–plasma interactions as a new field of research have received considerable attention for strong THz radiation generation[Reference Hamster, Sullivan, Gordon, White and Falcone6, Reference Leemans, Geddes, Faure, Tóth, van Tilborg, Schroeder, Esarey, Fubiani, Auerbach, Marcelis, Carnahan, Kaindl, Byrd and Martin22Reference Gopal, Herzer, Schmidt, Singh, Reinhard, Ziegler, Brömmel, Karmakar, Gibbon, Dillner, May, Meyer and Paulus26]. However, with increasing laser energies, the repetition rates and stability of laser systems are both decreased. The sensitivity of laser–plasma interactions to small laser fluctuations has motivated the development of true single-shot THz-diagnostic techniques.

3.1. Incoherent detection

Direct measurements based on thermal absorption are still by far the most convenient and widely used methods for the detection of broadband plasma-based THz sources. In order to obtain spectral information in one laser shot, multi-channel detection with bandpass or low-pass filters is preferred, as depicted in Figure 3. The detector array usually consists of a number of identical pyroelectric elements; or in the case of sufficient radiation flux, a pyroelectric camera can be considered as a receiver. Selectable beam splitters include silicon wafers and pellicles; coupling devices can be from a wide range of THz lenses and waveguides. By deliberate calibration of the split ratios, coupling coefficients, filter transmission curves and detector response curves, the essential spectral information of THz pulses can be obtained from one laser shot. This multi-channel THz-diagnostic system provides a convenient way to avoid shot-to-shot laser fluctuations, and offers the flexibility of selecting the frequency range of interest simply by changing the THz filters.

Figure 3. Schematic of a multi-channel THz-diagnostic system based on a direct detector array.

3.2. Coherent detection

THz pulses generated from ultrafast laser–plasma interactions are often characterized by pump and probe measurements. Traditional pump–probe schemes such as EO sampling, which relies on lock-in techniques, are very sensitive to laser noise (which can be quite substantial at low repetition rates) and low-frequency disturbance. For intense laser–plasma interactions, the low repetition rate and relatively large laser fluctuations have motivated the development of true single-shot EO-based measurements. In principle, the non-scanning schemes introduced in Section 2 can be used in a single-shot measurement if a reference beam be provided along with the probe beam in one laser shot for direct comparison[Reference Li, Li, Zhou, Wang, Du, Ding, Lin, Liu, Sheng, Peng, Chen, Ma, Lu, Wang, Wei and Zhang25]. In this section, two further single-shot techniques via temporal–spectral or temporal–spatial conversion will be introduced.

3.2.1. Single-shot interferometric scheme using frequency-domain holography

Spectral interferometry, which is also called Fourier-domain interferometry (FDI), is a well-known linear technique for phase and amplitude retrieval from an unknown field in the femtosecond domain[Reference Geindre, Audebert, Rebibo and Gauthier27]. Normally, multiple shots are required to obtain the complete temporal field, but the acquisition rate can be greatly increased by using linearly chirped probe pulses. The in-line interferometric scheme is considered unsuitable for true single-shot THz field diagnostics, since a reference shot has to be recorded as the background before performing the signal retrieval[Reference Kim, Yellampalle, Rodriguez, Averitt, Taylor and Glownia20]. We proposed a modified scheme using single-shot frequency-domain holography (FDH)[Reference Zhou, Liu, Li, Du, Li, Wang, Sheng, Chen, Ma, Lu, Dong and Zhang28], as depicted in Figure 4. The original femtosecond optical pulse is frequency chirped and time stretched to several picoseconds before splitting by a 50∕50 beam splitter (BS1). A fixed time delay ${\it\tau}$ is introduced to overlap the probe pulse within the duration of the THz pulse in the ZnTe crystal. The delayed and modulated probe pulse then interferes with the reference pulse in an imaging spectrometer, and the interferometric fringes are recorded on a CCD camera. The interference pattern, with a period proportional to $2{\it\pi}/{\it\tau}$, which contains information on the phase and amplitude modulation caused by the THz field, will then be analyzed to retrieve the THz waveform in a single shot. The temporal resolution of this technique relies on the retrieval algorithm. If the signal recovery is based on the linear chirped time–frequency relation, a temporal limit identical to that of the spectral encoding method will be given by the uncertainty principle[Reference Sun, Jiang and Zhang18]. The implementation of such a technique requires use of a high-resolution spectrometer.

Figure 4. Spectral interferometry scheme using twin-chirped-pulse FDH (quoted from Ref. [Reference Zhou, Liu, Li, Du, Li, Wang, Sheng, Chen, Ma, Lu, Dong and Zhang28]).

3.2.2. Single-shot THz-pulse characterization by dual echelon optics

In the 2000s, dual echelon optics was first employed to single-shot THz diagnostics[Reference Kim, Yellampalle, Taylor, Rodriguez and Glownia29]. The main idea is to divide the transverse profile of a probe beam by stair-step (echelon) structures into multiple delayed beamlets, which are then focused onto the same spot on an EO crystal to monitor the phase modulation caused by a copropagating THz pulse. The portions of probe beamlets with THz temporal information encoded on them are sequentially imaged onto different positions of a CCD camera for single-shot image processing. To maximize the number of sampling points within one laser shot, two orthogonally oriented echelons are used (rather than one) to fully subdivide the transverse probe beam profile, which has been pre-magnified in diameter. The so-called echelon optics technique, as depicted in Figure 5, which uses a collinear geometry with focused copropagating THz and probe beams, maintains a high detection sensitivity while providing high time resolution and a large time window. For a true single-shot measurement, a beam splitter (blue dot line in Figure 5) should be applied in the optical path to provide an additional reference pulse which bypasses the ZnTe crystal. The reference image is used for direct comparison with the signal image within one laser shot.

Figure 5. Schematic of single-shot THz measurement using dual echelon optics (quoted from Ref. [Reference Kim, Yellampalle, Taylor, Rodriguez and Glownia29]).

3.3. Infrared streak camera

Most electronic devices cannot be used for the direct measurement of fast transient processes in the picosecond region, with the exception of streak cameras. The spectral response of a streak camera is determined by the photocathode material, most of which are sensitive only to electromagnetic radiation of wavelengths shorter than $1.5~{\rm\mu}\text{m}$. By using gas-phase Rydberg atoms as the cathode, the measurable wavelength can be extended to the FIR region[Reference Drabbels and Noordam30]. Alternatively, by placing an EO crystal as a converter before an optical streak camera, THz pulses can be retrieved indirectly by measuring copropagating long probe pulses[Reference Jiang, Sun and Zhang31]. Comparing an atomic streak camera and an EO-based infrared streak camera, the former can provide information only on the time envelope of THz waves, while the latter is in principle a coherent measurement (but requires an additional reference shot for THz field diagnosis). The temporal resolution of infrared streak camera systems is not comparable to that which can be achieved by single-shot EO-based optical systems.

4. Summary and outlook

In this paper, we present an overview on the state-of-the-art techniques for single-shot THz diagnostics involved in intense laser–plasma interactions. For incoherent detection, direct detectors can be used with THz filters to provide general spectral information. Atomic infrared streak cameras, which operate in the infrared regime (typ. from 1 to $100~{\rm\mu}\text{m}$), are able to measure the intensity profile of THz pulses directly, with a time resolution of about one picosecond. For coherent detection, a THz field can be retrieved via single-shot EO-sampling techniques, either by time-to-space or time-to-spectrum conversion. Associated with specific applications, THz-diagnostic systems should be carefully designed according to the experimental and laser conditions. Practically speaking, high detection sensitivity is always the first concern, followed by high time resolution. Efforts toward higher-power THz sources and an improved understanding of THz radiation under extreme conditions, would promise continuing advances in single-shot THz technologies for intense laser–plasma interactions.

Acknowledgements

This work is supported by the National Basic Research Program of China (grant no. 2013CBA01501), the National Nature Science Foundation of China (grant nos. 11135012 and 11375262), and the Key Research Program of the Chinese Academy of Science (grant no. KGZD-EW-T05).

References

Ferguson, B. and Zhang, X.-C. Nat. Mater. 1, 26 (2002).Google Scholar
Siegel, P. H. IEEE Trans. Microwave Theory Technol. 50, 910 (2002).Google Scholar
Tzortzakis, S. Méchain, G. Patalano, G. André, Y.-B. Prade, B. Franco, M. Mysyrowicz, A. Munier, J.-M. Gheudin, M. Beaudin, G. and Encrenaz, P. Opt. Lett. 27, 1944 (2002).Google Scholar
Gaidis, M. C. Pickett, H. M. Smith, C. D. Smith, R. P. Martin, S. C. and Siegel, P. H. IEEE Trans. Microwave Theory Technol. 48, 733 (2000).Google Scholar
Dolan, G. J. Phillips, T. G. and Woody, D. P. Appl. Phys. Lett. 34, 347 (1979).Google Scholar
Hamster, H. Sullivan, A. Gordon, S. White, W. and Falcone, R. W. Phys. Rev. Lett. 71, 2725 (1993).Google Scholar
Cai, Y. Brener, I. Lopata, J. Wynn, J. Pfeiffer, L. Stark, J. B. Wu, Q. Zhang, X. C. and Federici, J. F. Appl. Phys. Lett. 73, 444 (1998).Google Scholar
Park, S.-G. Melloch, M. R. and Weiner, A. M. Appl. Phys. Lett. 73, 3184 (1998).Google Scholar
Kono, S. Tani, M. Gu, P. and Sakai, K. Appl. Phys. Lett. 77, 4104 (2001).Google Scholar
Kübler, C. Huber, R. and Leitenstorfer, A. Semicond. Sci. Technol. 20, S128 (2005).Google Scholar
Gallot, G. and Grischkowsky, D. J. Opt. Soc. Am. B 16, 1204 (1999).Google Scholar
Wu, Q. and Zhang, X.-C. Appl. Phys. Lett. 68, 1604 (1996).Google Scholar
Wu, Q. and Zhang, X.-C. Appl. Phys. Lett. 70, 1784 (1997).Google Scholar
van Tilborg, J. Schroeder, C. B. Filip, C. V. Tóth, Cs. Geddes, C. G. R. Fubiani, G. Huber, R. Kaindl, R. A. Esarey, E. and Leemans, W. P. Phys. Rev. Lett. 96, 014801 (2006).Google Scholar
Nahata, A. Weling, A. S. and Heinz, T. F. Appl. Phys. Lett. 69, 2321 (1996).Google Scholar
Jiang, Z. and Zhang, X.-C. Appl. Phys. Lett. 72, 1945 (1998).Google Scholar
Shan, J. Weling, A. S. Knoesel, E. Bartels, L. Bonn, M. Nahata, A. Reider, G. A. and Heinz, T. F. Opt. Lett. 25, 426 (2000).Google Scholar
Sun, F. G. Jiang, Z. and Zhang, X.-C. Appl. Phys. Lett. 73, 2233 (1998).Google Scholar
Jamison, S. P. Shen, J. MacLeod, A. M. Gillespie, W. A. and Jarpszynski, D. A. Opt. Lett. 28, 1710 (2003).Google Scholar
Kim, K. Y. Yellampalle, B. Rodriguez, G. Averitt, R. D. Taylor, A. J. and Glownia, J. H. Appl. Phys. Lett. 88, 041123 (2006).Google Scholar
Yellampalle, B. Kim, K. Y. Rodriguez, G. Glownia, J. H. and Taylor, A. J. Appl. Phys. Lett. 87, 211109 (2005).Google Scholar
Leemans, W. P. Geddes, C. G. R. Faure, J. Tóth, Cs. van Tilborg, J. Schroeder, C. B. Esarey, E. Fubiani, G. Auerbach, D. Marcelis, B. Carnahan, M. A. Kaindl, R. A. Byrd, J. and Martin, M. C. Phys. Rev. Lett. 91, 074802 (2003).Google Scholar
Sheng, Z.-M. Mima, K. Zhang, J. and Sanuki, H. Phys. Rev. Lett. 94, 095003 (2005).Google Scholar
Kim, K. Y. Taylor, A. J. Glownia, J. H. and Rodriguez, G. Nat. Photonics 2, 605 (2008).Google Scholar
Li, Y. T. Li, C. Zhou, M. L. Wang, W. M. Du, F. Ding, W. J. Lin, X. X. Liu, F. Sheng, Z. M. Peng, X. Y. Chen, L. M. Ma, J. L. Lu, X. Dong Wang, Z. H. Wei, Z. Y. and Zhang, J. Appl. Phys. Lett. 100, 254101 (2012).Google Scholar
Gopal, A. Herzer, S. Schmidt, A. Singh, P. Reinhard, A. Ziegler, W. Brömmel, D. Karmakar, A. Gibbon, P. Dillner, U. May, T. Meyer, H.-G. and Paulus, G. G. Phys. Rev. Lett. 111, 074802 (2013).Google Scholar
Geindre, J.-P. Audebert, P. Rebibo, S. and Gauthier, J.-C. Opt. Lett. 26, 1612 (2001).Google Scholar
Zhou, M. L. Liu, F. Li, C. Du, F. Li, Y.-T. Wang, W.-M. Sheng, Z.-M. Chen, L.-M. Ma, J.-L. Lu, X. Dong, Q.-L. and Zhang, J. Chinese Phys. Lett. 29, 015202 (2012).Google Scholar
Kim, K. Y. Yellampalle, B. Taylor, A. J. Rodriguez, G. and Glownia, J. H. Opt. Lett. 32, 1968 (2007).Google Scholar
Drabbels, M. and Noordam, L. D. Appl. Phys. Lett. 74, 1797 (1999).Google Scholar
Jiang, Z. Sun, F. G. and Zhang, X.-C. Opt. Lett. 24, 1245 (1999).Google Scholar
Figure 0

Figure 1. Schematic of (a) spectral encoding and (b) spatial encoding scheme.

Figure 1

Figure 2. Schematic of ‘single-shot’ cross-correlation scheme.

Figure 2

Figure 3. Schematic of a multi-channel THz-diagnostic system based on a direct detector array.

Figure 3

Figure 4. Spectral interferometry scheme using twin-chirped-pulse FDH (quoted from Ref. [28]).

Figure 4

Figure 5. Schematic of single-shot THz measurement using dual echelon optics (quoted from Ref. [29]).